摘要:The aim of this work is to analyze the function of natural teeth’s periodontal ligament and the effects of periodontal ligament on implants by the finite element method (FEM), when static functional loads occur. The finite element analysis models are established, which consist of fragment of mandible, natural teeth, periodontal ligament, and implant. Finite element software Abaqus is used to analyze the stress transfer in models with or without periodontal ligament, and mandibular first molar is implant. The implant obtained the maximum stress value of 87.71MPa when periodontal ligament of natural tooth is absent, but the maximum value reduces to 38.43 MPa with the action of periodontal ligament. It illustrates that periodontal ligament has significant effects on stress transfer. When the finite element model of single natural tooth or dentition with implant is generated, periodontal ligament should be taken into account.