标题:Estimating the whole-body effective dose and health risks as well as introducing a new easy method for eye lens dosimetry in interventional cardiology procedures
摘要:AbstractThis study aimed to introduce a new method for eye lens thermo-luminescent dosimetry and also estimate the dose associated with induced cancer risk due to the ionizing radiation exposure received by physicians and other staff cooperating in interventional cardiology (IC) procedures. The measurements were performed with six TLDs (thermoluminescent dosimeters): four TLDs for eye lens dosimetry (2 positioned on respiratory/surgical mask under the eye region as the new method; and 2 near the outside border of the eye as the common method) and two TLDs for whole-body dosimetry. Whole-body doses were used to calculate the cancer risks induced by IC procedures. The results of the new proposed method for eye lens dosimetry were similar to common TLD positioning (mean differences <5%) and mask displacement had no significant effect on eye dose measurement in our new method. Our proposed method for eye lens dosimetry is simpler and more comfortable compared to the common method and it can be used as an alternative method without using TLD holders to monitor lens dose for IC workers wearing masks during the procedure. The estimated excess cancer incidence risk induced by IC procedures was 29.58 ± 5.71 and 46.68 ± 7.77 (per 100000 individuals) for men and women, respectively.Graphical abstractDisplay Omitted