首页    期刊浏览 2024年10月06日 星期日
登录注册

文章基本信息

  • 标题:Design of a loading system for cyclic test on sutured organs
  • 本地全文:下载
  • 作者:Giulia Pascoletti ; Maria Chiara Pressanto ; Giovanni Putame
  • 期刊名称:MethodsX
  • 印刷版ISSN:2215-0161
  • 电子版ISSN:2215-0161
  • 出版年度:2020
  • 卷号:7
  • 页码:1-12
  • DOI:10.1016/j.mex.2020.100988
  • 语种:English
  • 出版社:Elsevier
  • 摘要:AbstractThe design of loading systems to test biologic samples is often challenging, due to shape variability and non-conventional loading set-ups. In addition to this, large economic investments would not be justified since the loading set up is usually designed for one single or for a limited range of applications.The object of this work is the development of a loading set-up finalised to on-site testing of sutures whose main function is applying a localised tensile load. The main challenges of this design process can be so summarized:•Applying cyclic tensile loads on the suture wire, mimicking the physiologic condition where both suture anchorage points have a certain compliance;•Designing a loading system as versatile as possible, in order to be able to accommodate organs with different geometries and sizes;•Keeping low both the complexity and costs of realization.All these considerations and the design calculi are here reported in detail, discussing the novelty of the system, and its main advantages.Graphical abstractDisplay Omitted
  • 关键词:Suture test;Suture distraction;Cyclic loads;Anchorage points migration