摘要:The closed flash point is an important indicator in the detection of organic heat carrier. In this paper, based on the Pensky-Martin flash point (closed) measurement method, the closed flash point of the organic heat carrier is measured with an automatic closed cup flash point meter. The factors that may affect the accuracy of the measurement result during the measurement process are analyzed, and the uncertainty is determined. Through the analysis of the established uncertainty component sources, the repeatability measurement of the organic heat carrier, the temperature sensor, the pressure sensor, and the numerical rounding based on the standard, the results of the closed flash point determination of the organic heat carrier are obtained. The synthetic uncertainty and the relative expanded uncertainty are evaluated to improve the reliability of the measurement results of the closed flash point of the organic heat carrier and provide a reference for the quality control of the organic heat carrier.