首页    期刊浏览 2025年06月30日 星期一
登录注册

文章基本信息

  • 标题:Comparing prediction accuracy between total keratometry and conventional keratometry in cataract surgery with refractive multifocal intraocular lens implantation
  • 本地全文:下载
  • 作者:Ho Seok Chung ; Jae Lim Chung ; Young Jun Kim
  • 期刊名称:Scientific Reports
  • 电子版ISSN:2045-2322
  • 出版年度:2021
  • 卷号:11
  • DOI:10.1038/s41598-021-98491-x
  • 语种:English
  • 出版社:Springer Nature
  • 摘要:We aimed to compare refractive outcomes between total keratometry using a swept-source optical biometer and conventional keratometry in cataract surgery with refractive multifocal intraocular lens (IOL) implantation. We included patients who underwent cataract surgery with refractive multifocal IOL implantation. The IOL power was calculated using conventional formulas (Haigis, SRK/T, Holladay 2, and Barrett Universal II) as well as a new formula (Barrett TK Universal II). The refractive mean error, mean absolute error, and median absolute error were compared, as were the proportions of eyes within ± 0.25 diopters (D), ± 0.50 D, and ± 1.00 D of prediction error. In total 543 eyes of 543 patients, the absolute prediction error of total keratometry was significantly higher than that of conventional keratometry using the SRK/T ( P  = 0.034) and Barrett Universal II ( P  = 0.003). The proportion of eyes within ± 0.50 D of the prediction error using the SRK/T and Barrett Universal II was also significantly higher when using conventional keratometry than total keratometry ( P  = 0.010 for SRK/T and P  = 0.005 for Barrett Universal II). Prediction accuracy of conventional keratometry was higher than that of total keratometry in cataract surgery with refractive multifocal IOL implantation.
国家哲学社会科学文献中心版权所有