首页    期刊浏览 2024年10月05日 星期六
登录注册

文章基本信息

  • 标题:Evaluation of the reliability of lower extremity alignment measurements using EOS imaging system while standing in an even weight-bearing posture
  • 本地全文:下载
  • 作者:Byung Woo Cho ; Tae-Ho Lee ; Sungjun Kim
  • 期刊名称:Scientific Reports
  • 电子版ISSN:2045-2322
  • 出版年度:2021
  • 卷号:11
  • DOI:10.1038/s41598-021-01646-z
  • 语种:English
  • 出版社:Springer Nature
  • 摘要:This study aimed to analyze the reproducibility and reliability of the alignment parameters measured using the EOS image system in both limbs while standing with an even weight-bearing posture. Overall, 104 lower extremities in 52 patients were analyzed retrospectively. The patients stood with an even load over both lower extremities then rotated 15° in both directions. Two EOS images were acquired and 104 pairs of lower extremities were compared according to the position of the indexed lower extremities. Then, the inter-observer reliability of the EOS system and the inter-modality reliability between EOS and computed tomography (CT) were evaluated. Femoro-tibial rotation (FTR) and tibial torsion demonstrated a significant difference between the anterior and posterior positions of the indexed lower extremity. In the inter-observer reliability analysis, all values except for FTR and tibial torsion demonstrated good or very good reliability. In the anterior position, FTR demonstrated moderate, and tibial torsion demonstrated poor reliability. In the posterior position, both FTR and tibial torsion demonstrated poor reliability. In the reliability analysis between the three-dimensional (3D) EOS model and 3D CT images, all measurements of the femur demonstrated very good reliability, but measurements of the tibia did not. For the coronal and sagittal alignment parameters measured by the EOS 3D system with rotated standing posture, except for the measurement including tibial torsion., there were no significant difference for either position of the indexed extremities with high agreement between the observers as well as with the CT 3D model.
国家哲学社会科学文献中心版权所有