  • 标题:Accuracy Assessment of the GlucoMen ® Day CGM System in Individuals with Type 1 Diabetes: A Pilot Study
  • 本地全文:下载
  • 作者:Daniel A. Hochfellner ; Amra Simic ; Marlene T. Taucher
  • 期刊名称:Biosensors
  • 电子版ISSN:2079-6374
  • 出版年度:2022
  • 卷号:12
  • 期号:2
  • DOI:10.3390/bios12020106
  • 语种:English
  • 出版社:MDPI Publishing
  • 摘要:The aim of this study was to evaluate the accuracy and usability of a novel continuous glucose monitoring (CGM) system designed for needle-free insertion and reduced environmental impact. We assessed the sensor performance of two GlucoMen ® Day CGM systems worn simultaneously by eight participants with type 1 diabetes. Self-monitoring of blood glucose (SMBG) was performed regularly over 14 days at home. Participants underwent two standardized, 5-h meal challenges at the research center with frequent plasma glucose (PG) measurements using a laboratory reference (YSI) instrument. When comparing CGM to PG, the overall mean absolute relative difference (MARD) was 9.7 [2.6–14.6]%. The overall MARD for CGM vs. SMBG was 13.1 [3.5–18.6]%. The consensus error grid (CEG) analysis showed 98% of both CGM/PG and CGM/SMBG pairs in the clinically acceptable zones A and B. The analysis confirmed that GlucoMen ® Day CGM meets the clinical requirements for state-of-the-art CGM. In addition, the needle-free insertion technology is well tolerated by users and reduces medical waste compared to conventional CGM systems.
  • 关键词:endiabetes technologyCGMaccuracytype 1 diabetessustainability