  • 标题:Variable Classifications of Glycemic Index Determined by Glucose Meters
  • 本地全文:下载
  • 作者:Meng-Hsueh Amanda Lin ; Ming-Chang Wu ; Jenshinn Lin
  • 期刊名称:Journal of Clinical Biochemistry and Nutrition
  • 印刷版ISSN:0912-0009
  • 电子版ISSN:1880-5086
  • 出版年度:2010
  • 卷号:47
  • 期号:1
  • 页码:45-52
  • DOI:10.3164/jcbn.10-08
  • 出版社:The Society for Free Radical Research Japan
  • 摘要:The study evaluated and compared the differences of glucose responses, incremental area under curve (IAUC), glycemic index (GI) and the classification of GI values between measured by biochemical analyzer (Fuji automatic biochemistry analyzer (FAA)) and three glucose meters: Accue Chek Advantage (AGM), BREEZE 2 (BGM), and Optimum Xceed (OGM). Ten healthy subjects were recruited for the study. The results showed OGM yield highest postprandial glucose responses of 119.6 ± 1.5, followed by FAA, 118.4 ± 1.2, BGM, 117.4 ± 1.4 and AGM, 112.6 ± 1.3 mg/dl respectively. FAA reached highest mean IAUC of 4156 ± 208 mg × min/dl, followed by OGM (3835 ± 270 mg × min/dl), BGM (3730 ± 241 mg × min/dl) and AGM (3394 ± 253 mg × min/dl). Among four methods, OGM produced highest mean GI value than FAA (87 ± 5) than FAA, followed by BGM and AGM (77 ± 1, 68 ± 4 and 63 ± 5, p <0.05). The results suggested that the AGM, BGM and OGM are more variable methods to determine IAUC, GI and rank GI value of food than FAA. The present result does not necessarily apply to other glucose meters. The performance of glucose meter to determine GI value of food should be evaluated and calibrated before use.
  • 关键词:glycemic index;brown rice;yogurt drink;mango;glucose meter