首页    期刊浏览 2024年11月23日 星期六
登录注册

文章基本信息

  • 标题:A Case of Minimally Invasive Aortic Valve Replacement Surgery Using a 19mm SJM TrifectaTM Bioprosthetic Valve
  • 本地全文:下载
  • 作者:Shunichi Kondo ; Tomohiro Takano ; Ken Nakamura
  • 期刊名称:循環制御
  • 印刷版ISSN:0389-1844
  • 出版年度:2012
  • 卷号:33
  • 期号:3
  • 页码:209-212
  • DOI:10.11312/ccm.33.209
  • 出版社:Japan Society of Circulation Control in Medicine
  • 摘要:A St. Jude Medical(SJM) Trifecta valve was developed as a new tissue valve by improving the previous tissue valve. The effective orifice area is wider than that of Epic tissue valve. The efficacy of the new valve has not been reported yet. Recently, we first performed aortic valve replacement(AVR) with the SJM Trifecta valve for aortic valve stenosis. A case 74-year-old female was admitted to our hospital with dyspnea and angina on exertion. She was diagnosed aortic valve stenosis. She underwent AVR with a 19 mm TrifectaTM tissue valve. According to the intra-operative direct simultaneous pressure measurement conducted after weaning of cardio pulmonary bypass, the peak pressure gradient of the prosthetic valve was 11 mmHg, the mean pressure gradient was 5 mmHg. According to the results of echocardiography conducted 10 days postoperatively, the peak pressure gradient of the prosthetic valve was 24.4 mmHg, the mean pressure gradient was 16.5 mmHg. She was discharged 18 days after surgery without complications. Implantation of a 19mm SJM TrifectaTM valve produced excellent result reflected by lower pressure gradient and absence of patient-prosthetic mismatch. In the future, the new valve is expected to be the optimum tissue valve for a narrow annulus.
  • 关键词:tissue valve ; MICS ; AVR ; Trifecta