首页    期刊浏览 2024年11月24日 星期日
登录注册

文章基本信息

  • 标题:Comparison of Dynamic Contour Tonometry and Goldmann Applanation Tonometry in Relation to Central Corneal Thickness
  • 本地全文:下载
  • 作者:Kim, Min Kyo ; Park, Si Yoon ; Kim, Chan Yun
  • 期刊名称:Journal of the Korean Ophthalmological Society
  • 印刷版ISSN:0378-6471
  • 出版年度:2015
  • 卷号:56
  • 期号:9
  • 页码:1392-1399
  • DOI:10.3341/jkos.2015.56.9.1392
  • 语种:Korean
  • 出版社:The Korean Ophthalmological Society
  • 摘要:Purpose

    In this study we compared the intraocular pressures (IOPs) measured using dynamic contour tonometry (DCT) and Goldmann applanation tonometry (GAT) and investigated the correlation between central corneal thickness (CCT) and IOP.

    Methods

    In a prospective study, 178 eyes of 91 subjects with glaucoma and glaucoma suspect were enrolled. IOP was measured using DCT and GAT and CCT was measured using ultrasound pachymetry. Each eye was classified into 1 of 3 groups according to their CCT: low CCT (Group A; CCT < 525 µm), normal CCT (Group B; 525 ≤ CCT < 561 µm), and high CCT (Group C; CCT ≥ 561 µm). In each group, we investigated the correlation between CCT and IOP measurement using GAT and DCT.

    Results

    A significant correlation was found between CCT and IOP measured using GAT ( p < 0.001), but not between CCT and IOP measured using DCT ( p = 0.108) in all patients. Subgroup analysis showed that CCT affected IOP measured with GAT only in Group A ( p = 0.027) and IOP measured with DCT was not affected by CCT in all 3 groups.

    Conclusions

    IOP measured using GAT and DCT were significantly affected by CCT, especially when the CCT was low. CCT may have no effect on DCT thus, IOP measured using DCT is more reliable than GAT in patients with thin corneas.

  • 关键词:Central corneal thickness; Dynamic contour tonometry; Glaucoma; Goldmann applanation tonometry
国家哲学社会科学文献中心版权所有