首页    期刊浏览 2024年07月08日 星期一
登录注册

文章基本信息

  • 标题:Biotelemetric Wireless Intracranial Pressure Monitoring: An In Vitro Study
  • 本地全文:下载
  • 作者:Mohammad H. Behfar ; Toni Björninen ; Elham Moradi
  • 期刊名称:International Journal of Antennas and Propagation
  • 印刷版ISSN:1687-5869
  • 电子版ISSN:1687-5877
  • 出版年度:2015
  • 卷号:2015
  • DOI:10.1155/2015/918698
  • 出版社:Hindawi Publishing Corporation
  • 摘要:Assessment of intracranial pressure (ICP) is of great importance in management of traumatic brain injuries (TBIs). The existing clinically established ICP measurement methods require catheter insertion in the cranial cavity. This increases the risk of infection and hemorrhage. Thus, noninvasive but accurate techniques are attractive. In this paper, we present two wireless, batteryless, and minimally invasive implantable sensors for continuous ICP monitoring. The implants comprise ultrathin (50 μm) flexible spiral coils connected in parallel to a capacitive microelectromechanical systems (MEMS) pressure sensor. The implantable sensors are inductively coupled to an external on-body reader antenna. The ICP variation can be detected wirelessly through measuring the reader antenna’s input impedance. This paper also proposes novel implant placement to improve the efficiency of the inductive link. In this study, the performance of the proposed telemetry system was evaluated in a hydrostatic pressure measurement setup. The impact of the human tissues on the inductive link was simulated using a 5 mm layer of pig skin. The results from the in vitro measurement proved the capability of our developed sensors to detect ICP variations ranging from 0 to 70 mmHg at 2.5 mmHg intervals.
国家哲学社会科学文献中心版权所有