首页    期刊浏览 2024年09月02日 星期一
登录注册

文章基本信息

  • 标题:Evaluation of Objective Accommodation Power in Different Age Groups Using an Auto Accommodation Refractometer
  • 本地全文:下载
  • 作者:Kang, Dong Wan ; Eom, Young Sub ; Rhim, Jay Won
  • 期刊名称:Journal of the Korean Ophthalmological Society
  • 印刷版ISSN:0378-6471
  • 出版年度:2016
  • 卷号:57
  • 期号:1
  • 页码:20-24
  • DOI:10.3341/jkos.2016.57.1.20
  • 语种:Korean
  • 出版社:The Korean Ophthalmological Society
  • 摘要:Purpose

    To evaluate the objective amplitude of accommodation in various age groups using an autorefractometer and to compare the results with subjective assessments on accommodation. Objective accommodative amplitude of artificial lenses in post-operative cataract patients was also measured.

    Methods

    In this study, 41 patients who visited Korea University Guro Hospital from July through August 2014 were categorized into 6 different age groups. Accommodative amplitude was measured using long-distance refraction at 5 m and short-distance refraction at 20 cm. In 20 patients, subjective amplitude of accommodation was measured using the minus lens technique and was compared with the measured objective accommodative amplitude. Accommodative amplitude was also measured in 8 patients who received cataract surgery.

    Results

    The mean accommodative ability in normal people was 2.38 D. The age groups of 10-19, 20-29, 30-39, 40-49, 50-59, and over 60 years showed accommodative amplitudes of 3.31 ± 0.99 D, 3.25 ± 0.34 D, 3.78 ± 0.68 D, 1.00 ± 0.60 D, 0.22 ± 0.23 D, 0.00 ± 0.00 D, respectively. The results showed a rapid decrease in accommodative ability of patients older than 40 years and demonstrated a correlation between objective and subjective amplitude of accommodation (Pearson's correlation coefficient 0.838). There was a significant difference between objective and subjective accommodative amplitudes in artificial lenses ( p = 0.015), with average objective and subjective measures of 0.10 ± 0.23 D and 0.59 ± 0.35 D, respectively.

    Conclusions

    Measuring amplitude of accommodation using an autorefractometer can be useful in assessing results of presbyopia treatment.

  • 关键词:Accommodation; Autorefractometer; Pseudoaccommodation
国家哲学社会科学文献中心版权所有