首页    期刊浏览 2024年07月08日 星期一
登录注册

文章基本信息

  • 标题:Left Atriotomy Entrapment of Pulmonary Artery Catheter during Mitral Valve Replacement -A case report-
  • 本地全文:下载
  • 作者:Shim, Jae Kwang ; Na, Sungwon ; Kim, Hong Sun
  • 期刊名称:Korean Journal of Anesthesiology
  • 印刷版ISSN:2005-6419
  • 出版年度:2006
  • 卷号:50
  • 期号:3
  • 页码:315-318
  • DOI:10.4097/kjae.2006.50.3.315
  • 语种:English
  • 出版社:The Korean Society of Anesthesiologists,
  • 摘要:

    A 51-yr-old man underwent mitral valve replacement and tricuspid valve repair due to mitral and tricuspid regurgitation under cardiopulmonary bypass. The pulmonary artery (PA) catheter was inserted easily via right internal jugular vein and functioned well until the end of surgery. The surgery was uneventful and patient's hemodynamics were stable both in the operating room and intensive care unit. All PA catheter functions were normal, but the balloon rupture was suspected because it was unable to obtain pulmonary capillary wedge pressure. On attempted removal of the PA catheter at the next day, considerable resistance was encountered and pulmonary artery tracing was dampened. A followed-up chest X-ray revealed the catheter to be in the right pulmonary artery without evidence of knotting. Suture related entrapment was suspected, and on reoperation, PA catheter was found to be sutured at the left atriotomy site. The PA catheter was removed under cardiopulmonary bypass and revealed a hole between proximal port and thermal filament.

  • 关键词:complication; Entrapment; Pulmonary artery catheter; surgical removal
国家哲学社会科学文献中心版权所有