首页    期刊浏览 2024年07月08日 星期一
登录注册

文章基本信息

  • 标题:Assessment of Blood Glucose Values Measured by Blood Gas Analyzer or Portable Glucometer
  • 本地全文:下载
  • 作者:Park, Hue Jung ; Park, Chul Soo ; Park, Chong Min
  • 期刊名称:Korean Journal of Anesthesiology
  • 印刷版ISSN:2005-6419
  • 出版年度:2006
  • 卷号:50
  • 期号:5
  • 页码:506-510
  • DOI:10.4097/kjae.2006.50.5.506
  • 语种:English
  • 出版社:The Korean Society of Anesthesiologists,
  • 摘要:

    BACKGROUND: A portable glucometer is commonly used to immediately check the blood glucose level. In the anesthetic field, some blood gas analyzers can also give a rapid indication of the blood sugar level but the accuracy is unknown. Therefore, this study assessed the accuracy of the blood glucose values measured by either a blood gas analyzer or portable glucometer.

    METHODS: Venous blood from diabetic patients was used to measure the glucose level with either a blood gas analyzer or a portable glucometer. The difference and 5% deviation from reference values was analyzed. These values were also assessed using a Bland-Altman plot and clinical significance was examined using a Clarke error grid.

    RESULTS: The differences from the reference values were smaller using the blood gas analyzer (1.3 ± 7.8 mg/dl) than using the portable glucometer (-5.1 ± 16.7 mg/dl)(P < 0.01). 73.4% of the values measured by the blood gas analyzer and 40.0% of those measured by the portable glucometer were within 5% of the reference value. The 95% limits of agreement in the difference ranged from -14.3 to 16.9 in the blood gas analyzer and -38.5 to 28.2 in the portable glucometer. Error grid analysis showed that 100% of the values measured by the blood gas analyzer were located in zone A. When locating the values measured using the portable glucometer, 95.6% were located in zone A, and the remaining 4.4% are located in zone B.

    CONCLUSIONS: The blood gas analyzer measures the blood glucose more accurately than the portable glucometer. However, the blood glucose values measured by the portable glucometer are clinically acceptable.

  • 关键词:bland altman analysis; blood gas analyzer; error grid analysis; Glucose; portable glucometer
国家哲学社会科学文献中心版权所有