首页    期刊浏览 2024年07月08日 星期一
登录注册

文章基本信息

  • 标题:Hypoxia from Erroneous Connection of a Nitrogen Tank for an Oxygen Tank: A case report
  • 作者:Kang, Seung Gwan ; Ko, Seong Hoon ; Lee, Sang Kyi
  • 期刊名称:Korean Journal of Anesthesiology
  • 印刷版ISSN:2005-6419
  • 出版年度:1999
  • 卷号:36
  • 期号:2
  • 页码:370-373
  • DOI:10.4097/kjae.1999.36.2.370
  • 语种:Korean
  • 出版社:The Korean Society of Anesthesiologists,
  • 摘要:

    We present a case of hypoxia which occurred during the onset of general anesthesia in a small hospital. It was found that one of the oxygen tank which formed the central pipeline gas supply had been erroneously replaced by a nitrogen tank. Lack of strict observance of Compressed Gas Supply Standards by the gas supplier and the hospital personnel allowed it. We also emphasize that the oxygen analyzer should be counted as an essential monitor in every anesthesia. Oxygen analyzer detects the supply of intraoperative hypoxic gas admixture promptly and effectively.

  • 关键词:Hypoxia; Monitoring, oxygen analyzer; Oxygen, delivery system
Loading...
联系我们|关于我们|网站声明
国家哲学社会科学文献中心版权所有