首页    期刊浏览 2024年11月27日 星期三
登录注册

文章基本信息

  • 标题:Comparison of the efficacy of disinfectants in automated endoscope reprocessors for colonoscopes: tertiary amine compound (Sencron2®) versus ortho-phthalaldehyde (Cidex®OPA)
  • 本地全文:下载
  • 作者:Seo, Hyun Il ; Lee, Dae Sung ; Yoon, Eun Mi
  • 期刊名称:Intestinal Research
  • 印刷版ISSN:1598-9100
  • 电子版ISSN:2288-1956
  • 出版年度:2016
  • 卷号:14
  • 期号:2
  • 页码:178-182
  • DOI:10.5217/ir.2016.14.2.178
  • 语种:English
  • 出版社:Korean Association for the Study of Intestinal Diseases
  • 摘要:Background/Aims

    To prevent the transmission of pathogens by endoscopes, following established reprocessing guidelines is critical. An ideal reprocessing step is simple, fast, and inexpensive. Here, we evaluated and compared the efficacy and safety of two disinfectants, a tertiary amine compound (TAC) and ortho-phthalaldehyde (OPA).

    Methods

    A total of 100 colonoscopes were randomly reprocessed using two same automated endoscope reprocessors, according to disinfectant. The exposure time was 10 minutes for 0.55% OPA (Cidex® OPA, Johnson & Johnson) and 5 minutes for 4% TAC (Sencron2®, Bab Gencel Pharma & Chemical Ind. Co.). Three culture samples were obtained from each colonoscope after reprocessing.

    Results

    A total of nine samples were positive among the 300 culture samples. The positive culture rate was not statistically different between the two groups (4% for OPA and 2% for TAC, P =0.501). There were no incidents related to safety during the study period.

    Conclusions

    TAC was non-inferior in terms of reprocessing efficacy to OPA and was safe to use. Therefore, TAC seems to be a good alternative disinfectant with a relatively short exposure time and is also less expensive than OPA.

  • 关键词:Colonoscopes; Cost; Disinfectants; Endoscopes
国家哲学社会科学文献中心版权所有