首页    期刊浏览 2024年07月05日 星期五
登录注册

文章基本信息

  • 标题:Adverse Effect of the Absorbable Rods in Treatment of the Radial Head
  • 本地全文:下载
  • 作者:Lee, Weon Ik ; Chang, Jun Dong ; Choi, Soo Joong
  • 期刊名称:Journal of the Korean Fracture Society
  • 印刷版ISSN:1225-1682
  • 出版年度:2002
  • 卷号:15
  • 期号:3
  • 页码:414-420
  • DOI:10.12671/jksf.2002.15.3.414
  • 语种:Korean
  • 出版社:The Korean Fracture Society
  • 摘要:

    PURPOSE: We report complications occurred from 6 patients among 14 patients who received the operation for their radial head and neck fractures by using the absorbable rod made by poly-glycolic acid(PGA). MATERIALS AND METHODS: We analyze the postoperative results of 14 patients who recieved fixation by absorbable rod for the radial head and neck fractures from March 1991 to March 2000. All of the fractures were are reducible and modified Mason 's type II. RESULTS: After average 15 months follow up, flexion contracture was average 20 degrees and full flexion was average 130 degrees. Complications were occurred in 6 cases. Osteolysis was occurred in 3 cases and in 2 cases among theses 3 cases, radial head excision was performed. Synovitis was occured in other 3 cases and in one case joint fluid was drainaged from operation wound for 2 weeks and in other 2 cases, synovitis was progressed to arthritis. CONCLUSION: The absorbable rod made of PGA in radial head and neck fracture have relatively high rate of adverse tissue responses. So surgeon should consider adverse tissue response of PGA. Development of more biocompatible absorbable and slow degrading material should be needed.

  • 关键词:radius; Head and neck fracture; Absorbable rod; Complication