  • 标题:The comparison among Mechanical, Electromyographic and Accelerographic responses during recovery from vecuronium induced neuromuscular blockade
  • 作者:Son, Hee Jeong ; Lee, Ji Heui ; Park, Sun Woo
  • 期刊名称:Korean Journal of Anesthesiology
  • 印刷版ISSN:2005-6419
  • 出版年度:1993
  • 卷号:26
  • 期号:5
  • 页码:910-918
  • DOI:10.4097/kjae.1993.26.5.910
  • 语种:Korean
  • 出版社:The Korean Society of Anesthesiologists,
  • 摘要:

    As the use of muscle relaxants routinize in the clinical general anesthesia, the proper estimation of neuromuscular blockade and recovery is needed for the more safe anesthetic management. Three recording methods of measurement in recovery are available; measurement of the evoked mechanical responses (mechanography), measurement of the evoked electrical responses (electromyography), and measurement of the acceleration of the muscle responses, The purpose of this study was to evaluate the correlations among mechanomyographic, electromyographic and accelerographic responses during recovery from vecuronium induced neuromuscular blockade. 15 ASA class 1 adult patients were studied. Measurements were as follows: 1) Recovery time of T(1)/Tc, of 25%, 50%, 75%, 90% and 100% in each device. 2) Recovery index in each device 3) The values of T(1)>Tc, in EMG and ACC corresponding ta that of MMG. 4) The values of TOF ratio in EMG and ACC corresponding to that of MMG. The results were as follows: 1) Recovery time of T(1).

  • 关键词:vecuronium; Mechanomyography; Electromyography; Accelerograph