To assess the degree of agreement of two rotating Scheimpflug cameras, Galilei G6 and Pentacam HR, in measuring corneal refractive power (K), anterior chamber depth (ACD), and central corneal thickness (CCT).
MethodsMeasurement agreement was assessed in 40 eyes of 40 outpatients at our hospital. Measurements of anterior and posterior corneal refractive power (K), ACD, and CCT were compared between the Galilei G6 and Pentacam HR.
ResultsFor Galilei G6 (4 mm), Pentacam HR (3 mm) and Pentacam HR (4 mm), the anterior corneal refractive powers (K) were 44.35 ± 1.38 D, 44.09 ± 1.32 D, and 44.12 ± 1.35 D, respectively, and the posterior corneal refractive powers (K) were 6.39 ± 0.23 D, 6.45 ± 0.23 D, 6.45 ± 0.22 D. The differences in the results were statistically significant. The average ACD measurements using Galilei G6 and Pentacam HR were 3.26 ± 0.42 mm and 3.17 ± 0.42 mm, respectively, and the average CCT measurements were 556.65 ± 30.12 µm and 553.78 ± 29.42 µm. The differences in the measurements were statistically significant. In addition, ACD 95% limits of agreement (LoA) between Galilei G6 and Pentacam HR were in the range of -0.14~0.32 mm, and CCT 95% LoA were in the range of -12.54~18.29 µm.
ConclusionsThere were significant differences in measurements of anterior and posterior corneal refractive power (K), ACD, and CCT between the two cameras. Agreement analysis suggests that Galilei G6 and Pentacam HR should not be used interchangeably.