Temperature fields, exchanges and deformations of a precise length comparator microscope/Precizinio ilgio komparatoriaus mikroskopo temperaturiniai laukai ir deformacijos.
Barauskas, R. ; Kasparaitis, A. ; Kausinis, S. 等
1. Introduction
Precision of determining the real position of the calibrated scale
depends highly on microscope stability [1]. To increase resolution power
and sensitivity of a microscope, optical schema and structure of its
precise length comparators are constantly improved by minimizing both
the impact of external disturbances on repeated measurements and
microscope drifts, especially those circumstanced by temperature
fluctuations.
Thermal expansion is a significant error source in measuring
systems which use lasers. Temperature gradients in optical and
mechanical components emerge due to the environmental temperature.
Stabilization of the instrument temperature at a constant environmental
temperature may last for several hours. When calibrating scales under
dynamic conditions, particularly at the fluctuating temperature in the
environment, the instrument and ambient temperatures never achieve a
good equilibrium. Instability of measurements lies mainly in
unhomogeneity of temperature deformations. For this reason great
attention is being given to modelling and experimental research into the
properties of length calibration microscopes [2-4].
To reduce temperature errors of precise microscope structures both
segmental structure and materials having a low thermal expansion
coefficient are used. 2D and 3D modelling and experimental research into
the prototype structures have indicated that sensitivity of such
structure to the temperature impact is considerably lower than that of
homogeneous structure since gaps between segments increase thermal
resistance with an insignificant decrease in stiffness of the structure.
Elaborating the structure of an investigated comparator, a modern
precise segmental microscope containing CCD camera and relay objectives
has been developed, FE model has been made, modelling of microscope
temperature deformations has been carried out and variation of
temperature deflections in time and in the microscope CCD camera
environment has been experimentally studied.
2. Problem formulation and solution methods
Precise length comparator (PLC) is a complex mechanical system, the
proper operation of which requires precision of geometrical dimensions
[5-7]. To comply with these requirements the orders of gradient
fractions of temperature fluctuations caused by temperature deformations
are of great importance. Therefore the impact of contingent
environmental temperature fluctuations on PLC geometry changes is to be
accurately estimated. Even with maximum elimination of mechanical
disturbances influence [8-10] and assurance of constant environmental
temperature security, certain temperature disturbances and the
deformations accompanying them are unavoidable due to the heat flows
spread by PLC electrical equipment. Though such deformations are
unavoidable, nevertheless proper structural solutions make it possible
either to reduce their impact on the measuring process or to develop
shading circuits if the extent of deformations can be in advance
estimated, or to recalculate the measurement results removing systematic
errors [11-13].
One of precarious temperature disturbances is the heat spread by
the measuring microscope CCD camera. As the steady-state temperature
under the operating conditions is known, the process can be calculated
by using the finite element simulation and the temperature values can be
found at all points of the microscope structure. Having the temperature
values obtained over the microscope structure, the displacements due to
thermal expansion can be calculated at all points of the structure.
The analysis scheme is as follows [6, 11]: calculating the static
or transient thermal field according to the assigned temperature values
at some structure points (of CCD camera surface) and in accordance with
the known conditions of convective heat exchange between the structure
and the ambient air, the equation of the structure heat balance reads as
follows
[C]{[??]} + [[K.sub.Th]]{T} = {[S.sub.[infinity]]} (1)
where [C] is matrix of thermal capacity, {[??]} is vector with the
time derivative of the nodal temperatures, [[K.sub.Th]] is matrix of
thermal conductivity, {T} is nodal temperatures vector,
{[S.sub.[infinity]]} is nodal vector of heat sources of the element
determined by the heat exchange over the surface of the body.
In the ANSYS system the heat exchange process is described by
element SOLID70. The solution presents the nodal temperature values,
which are further used as loads in the problem of thermal expansion of
the structure as:
[K]{U} = {Q} (2)
where [K] is stiffness matrix of the element, {U} is nodal
displacement matrix, {Q} is vector of nodal forces determined by
temperature loads.
The calculation scheme may be applied to the study of an impact of
environment temperature fluctuations on the other PLC elements, e.g. a
line scale, geometry changes. The finite element model of the structure
has been set up, in which the temperatures of the structure and the
ambient air could be calculated. The model is based on the coupling of
the following physical phenomena:
--heat transfer by the ambient air due to its thermal conductivity;
--convective heat transfer (due to the air fluctuation);
--heat exchange between the air and PLC structure;
--heat transfer by PLC structure due to its thermal conductivity;
--formation of deformations in PLC structure due to the
nonhomogenous thermal field generated in it.
[FIGURE 1 OMITTED]
In the computational model phenomena 1-4 have been described by
means of ANSYS (FLOTRAN) element FLUID142. The element can be used under
two different conditions:
--liquid (gas) dynamics described by the continuity equation, the
advection-diffusion equation and the ideal gas state equation;
--thermal conductivity in the solid described by the thermal
conductivity equation.
3. Model of microscope temperature deformations and calculation
results
The model of calculation of PLC microscope temperature deformations
is used to analyze two different structures: (Fig. 1, a), when CCD
camera is fixed onto the microscope side and (Fig. 1, b), when CCD
camera is fixed onto the axis of microscope symmetry
In both cases the temperatures in the zone of the camera fixture
are considered to be measured and known. The microscope is fixed on the
bracket by means of which its position in the space is adjusted. The
temperature at a certain distance from the bracket is assumed to be
equal to that of the structure. It is logical to assume that in the rest
of the structure surfaces the convectional heat exchanges with the
ambient air are proceeding. The value of thermal conductivity of
microscope material is assumed to be [[lambda].sub.m] = 54 W/mK, and
that of the camera and objective washers - [[lambda].sub.o] = 27 W/mK.
Convection coefficient over the surface is assumed to be everywhere -
[[beta].sub.c] = 20 W/[m.sup.2] K, and thermal expansion coefficient -
[[alpha].sub.t.exp] = 65 x [10.sup.-5].
[FIGURE 2 OMITTED]
Calculation results when CCD camera fixed on a microscope side
presented in Fig. 2, a and b. It should be noted that the temperature of
microscope beneath the bracket fixture rises insignificantly, and there
should be a good microscope bracket thermal contact (it is assumed in
the model).
Deviations are the magnitudes explained in Fig. 1, c. It is angle
[alpha] of the deformed axis with the vertical, as well as displacement
of cross-section center dz along the axis, as well as side displacement
ds. All three deviation types are observed: axis declination angle
[alpha], defocusing deviation dz and eccentric deviation ds .
[FIGURE 3 OMITTED]
To evaluate distribution of displacements not only visually but in
the context of the system functionality, deviation values of microscope
tube cross-sections position and shape in different cuts along the axis
are presented in Fig. 3.
[FIGURE 4 OMITTED]
In the structure when CCD camera is fixed on the symmetry axis,
axial deviations dominate which may affect only the quality of CCD ray
focusing, while less affecting its projection in the line scale. The
axis course remains unchanged.
4. Temperature fields in the microscope camera environment
4.1. Experimental research into temperature gradients in the space
In order to determine distribution of the temperature fields around
the microscope camera, temperature sensors were arranged in the way that
constant temperature curves (isotherms) could be represented in certain
cuts. To measure the temperature around the microscope camera, 21
sensors were used: four sensors were glued onto the microscope frame,
one--onto the flashbulb, one onto the camera and one onto the camera
setting device, the rest of them were layed out in the air.
The measurement process was divided into three phases:
--cut-off conditions, when the equipment is switched off for a long
time and acquires the ambient temperature;
--warming up, when the camera is switched on and warms up, but the
flashbulb does not operate;
--operating conditions, when the thermal balance between the camera
and the ambient air flux settles down.
Measurements have been performed three times under similar
conditions. Results of the first phase evaluated in terms of the
averages of three measurements indicate that the microscope temperature
is 0.15[degrees]C lower than that of the ambient air. This difference in
temperatures has emerged because before the experimental measuring of
the temperature fields the premises air-conditioner was switched off.
Temperatures around the microscope are uniformly distributed. Microscope
is slowly warming-up achieving the air flux temperature. In half an hour
its temperature increases by 0.04[degrees]C. A similar change is noticed
in the read-outs of sensors located closer to the microscope. The
closely spaced temperature field, Fig. 4, a, repeats the outline of the
microscope and camera frames. The shape of the farther field is cast
off.
When the camera is switched on, all sensors are recording heat.
Temperature dispersion rises. The temperature of sensors fixed onto the
frames of microscope and camera uniformly rises. Its least change occurs
in the sensors glued on the microscope frame farther from the camera.
The temperatures of the sensors which are closer to the camera rise a
little faster. A temperature change of the sensor fixed onto a flashbulb
is slight. The fastest rise in the temperature is indicated by the
sensor fixed onto a camera mandrel. It is evident that the camera
achieves thermal equilibrium in about 15 minutes.
Under operating conditions heat is withdrawn from the camera by
both the ambient air flux and microscope frame. The camera mandrel
disperses heat exceptionally well due to a big ratio of its surface area
to mass. Camera emits heat mostly through its couplings with cables on
the camera surface intended for data reading checkup and transmission.
The camera frame top warms up to 27.5[degrees]C. At a spot where camera
is connected to the bracket the temperature is only 24.7[degrees]C,
while that of camera sides is some fractions of a degree lower than on
the top. The bracket edge warms up to 22.5[degrees]C and after half an
hour of operation the sensors fixed onto microscope do not indicate any
difference. The temperature fields of horizontal cuts differ very little
from each other, whereas when comparing with the same cuts at a switched
off camera the distant field is drifting off.
In a vertical cut (Fig. 4) the visible field repeats the camera and
microscope outline. Its temperature from the camera to the top changes
according to the law which is approximate to the exponential one.
Conclusion may be drawn that the camera frame emits heat to the air
through both a holder and a mandrel--to microscope. Though the
microscope temperature change at the camera is negligible and no change
is recorded on the farther frame points, the camera emits the thermal
flux which affects temperature deformations of a microscope and a
calibrated scale.
4.2. Temperature gradients depending on time
To determine the change of temperature errors the contact
temperature measurements have been made on the top of the microscope
camera, on the camera setting device, on the camera flashbulb and the
air was measured at the line scale. Distribution of temperature sensors
is shown in Fig. 5. Temperature measurements have been made at some time
before scale calibration and during three calibrations, during the whole
measuring period the microscope camera was switched on.
[FIGURE 5 OMITTED]
The obtained results indicate that the temperature on the camera
top before calibration fluctuated by 0.2[degrees]C within the range of
27.4[degrees]C and 27.6[degrees]C (Fig. 5). During the first calibration
the temperature rose up to 27.8[degrees]C, and during the following two
calibrations it fluctuated within 27.7[degrees]C and 27.9[degrees]C.
This temperature fluctuation is stimulated by Peltier element present in
the camera which ensures its cooling.
Temperature change of the camera flashbulb frame is also clear.
Before calibration the temperature was stable and close to that of the
ambient air. From the start of the first calibration the temperature of
a flashbulb frame was continually increasing and at the end of the third
calibration the temperature change from the initial value reached
+0.6[degrees]C.
The temperature change on the camera setting device is not so much
significant as that on its top or on the flashbulb frame, nevertheless
the above mentioned factors have an influence on the temperature drift
up to 0.2[degrees]C (Fig. 6). This temperature drift affects temperature
deformations of the camera setting device and influences the general
uncertainty budget.
[FIGURE 6 OMITTED]
Before calibration and during three calibrations the temperature of
the air around the calibrated scale was being measured. During that
measurement time the air temperature was falling down within a certain
interval. This fall is influenced by a number of factors such as sources
of the increased amount of heat in the calibrated scale environment,
instability of air fluxes in the calibration laboratory, etc. All these
factors have an impact on the calibration results dispersion due to
thermal deformations.
5. Conclusions
1. CCD camera of the precise length comparator microscope for
determining the dash position of length measures is a source of
temperature calibration errors having an essential impact on calibration
process precision.
2. As a result of a camera direct temperature impact by means of
heat exchange the microscope frame thermally deforms yielding parasite
scale picture displacements formed in the microscope coming out as
calibration error constituents. Due to this temperature impact the
elements connecting the microscope calibrator to the laser
interferometer and also coming out as calibration error constituents
undergo thermal deformation. An additional error springs up due to both
a picture defocusing and a decrease in its distinctness.
3. Thermal emission of the camera violates standard temperature
conditions of calibration space of thermo-constant premises and brings
about temperature deformations of the measure being calibrated and the
comparator elements, thus also causing bigger calibration errors.
4. Calibration error caused by a thermal CCD camera impact under
steady-state calibration conditions is of a random character and in
real-time it cannot be compensated by mathematical methods.
5. In an investigated structure an estimate of the dispersion field
of this error at the 95% reliance interval is +/-0.23 [micro]m.
6. To minimize this calibration error sand it is expedient to use
the cameras of lower power, symmetric structures made of low thermal
expansion materials and to isolate CCD cameras thermally.
Received February 02, 2011
Accepted June 15, 2011
References
[1.] Barakauskas, A.; Kasparaitis, A.; Kausinis, S.; Lazdinas, R.
2009. Analysis of dynamic method of line scales detection, Solid State
Phenomena 147-149: 576-581
[2.] Bryan, J.B. 1990. International status of thermal error
research, Annals of the CIRP 39 (2): 645-656.
[3.] Yang, H. 2002. Dynamic Modeling for Machine Tool Thermal Error
Compensation. Ph.D. Dissertation, the University of Michigan.
[4.] Yang, H.; Ni, J. 2005. Dynamic neural network modeling for
nonlinear, nonstationary machine tool thermally induced error,
International Journal of Machine Tools and Manufacture 45(4-5): 455-465.
[5.] Koning R.; Flugge J.; Bosse H. 2007. A method for the In-Situ
determination of Abbe errors and their correction, Measurement Science
Technologies 18: 476-481
[6.] Jakstas A.; Kausinis S.; Barauskas R.; Barakauskas A.,
Kasparaitis A. 2009. Refinements of precision line scale calibration
system, In Proceedings 14th International Conference Mechanika, Kaunas
University of Technology, Kaunas, 137-140.
[7.] Kausinis S.; Barakauskas A.; Barauskas R.; Jakstas A.;
Kasparaitis A. 2008. Investigation of vibrations of line scale
calibration systems, Journal of Vibroengineering 10(1): 79-89.
[8.] Kilikevicius, A.; Jurevicius, M.; Berba, M. 2010. Research of
dynamics of a vibration isolation platform, Journal of Vibroengineering,
12(3): 361-367.
[9.] Brucas, D.; Giniotis V. 2009. Analysis of calibration data of
position of circular scale strokes, Journal of Vibroengineering, 11(1):
177-184.
[10.] Kilikevicius, A.; Petraska, A.; Juraitis S. 2009. Measurement
errors of comparator on carriage vibrations, Journal of Vibroengineering
11(4): 658-664.
[11.] Jakstas A.; Kausinis S.; Barauskas R.; Barakauskas A.;
Kasparaitis A. 2008. Software based control techniques for precision
line scale calibration, In Proceedings of the 11th Biennial Baltic
Electronics Conference, 2008 October 06-08, Tallinn University of
Technology, 5(8): 223-226.
[12.] Brucas, D.; Giniotis, V.; Augustinavicius, G.; Stepanoviene,
J. 2010. Calibration of the multiangular prism (polygon), Mechanika
4(84): 62-66.
[13.] Barakauskas, A.; Kasparaitis, A.; Kausinis, S.; Lazdinas, R.
2009. Analysis of estimation and compensation of angular errors of
linear motion, Mechanika 5(79): 52-58.
R. Barauskas, Kaunas University of Technology, Studentu 50, 51368
Kaunas, Lithuania, E-mail: rimantas.barauskas@ktu.lt
A. Kasparaitis, Vilnius Gediminas Technical University,
Basanaviciaus 28, 03224 Vilnius, Lithuania, E-mail:
a.kasparaitis@precizika.lt
S. Kausinis, Kaunas University of Technology, A. Mickeviciaus 37,
44244 Kaunas, Lithuania, E-mail: saulius.kausinis@ktu.lt
R. Lazdinas, Vilnius Gediminas Technical University, Basanaviciaus
28, 03224 Vilnius, Lithuania, E-mail: r.lazdinas@gmail.com