Arrangement for vertical angle calibration of geodetic instruments/Irenginys geodeziniu prietaisu vertikaliesiems kampams kalibruoti.
Giniotis, V. ; Brucas, D. ; Siaudinyte, L. 等
1. Introduction
Many opto-electronic digital instruments, such as rotary encoders,
theodolites, total stations, laser trackers, etc. are used in machine
engineering and instrumentation, geodesy, surveying, robotics and other
branches of industry. Most of optical--electronic geodetic measuring
instruments consist, among the other elements, of the circular scales
and angular transducers for angle determination in two perpendicular
planes--horizontal and vertical. Accuracy of the instrument mostly
depends on the accuracy of these angle measuring instruments. Metrology
of the optical instruments for horizontal and vertical angle
measurements has some specific features and needs specific arrangements
for its calibration, especially this concerns for vertical angle
calibration [1, 2]. Here we present a new approach to methodology and
means that can be used for the vertical angle calibration of geodetic
instruments.
Most of geodetic instruments have two angle reading devices
installed--for horizontal and vertical angle measurement. A number of
methods of calibration of the horizontal angle measurements are
implemented in practice, their origin comes from the circular scales and
rotary encoders calibration [3-5]. Here we analyse a proposal for
arrangement to create the reference standard for angle measurement
suitable for vertical angle calibration purposes in laboratory
environment. Calibration and testing of the geodetic angle measuring
instruments has always been a serious problem and if calibration of the
horizontal angle measurements could be quite efficiently accomplished
using standard precise turn tables (quite widely implemented in
metrology and industry), the calibration of vertical angle measures
required some special instrumentation.
Calibration of the accuracy of vertical angle measurements by the
geodetic instrument is usually performed using a special test bench
composed of autocollimators attached at the different vertical angles to
the calibrated instrument (Fig. 1, a). In this case the test bench is
extremely bulky and able to measure only very limited number of vertical
angles [2, 3]. A new design of the object was implementation of the
precise angle encoder as the reference angle standard (Fig. 1, b). In
this case it was possible to create unlimited number of reference angle
values although the equipment is extremely expensive.
The field testing [1] of accuracy of vertical angle measurement of
geodetic instrument is arranged by geometric angle consisting from
horizontal and vertical lines previously measured with high accuracy. As
they are usually of long distances, it is obvious that assurance of high
accuracy of these lines remains a complicated problem both in accuracy
and technical aspects. Its use is not simple and depends on various
factors, as temperature changes of environment atmosphere,
precipitation, summer or winter atmosphere, etc.
[FIGURE 1 OMITTED]
Here we present some development of the new arrangement for
vertical angle calibration of geodetic instruments [4--6] based on the
trigonometric method (angle determination by means of linear
measurements) and requires minimal amount of complicated
instrumentation.
The reference measure used for the calibration of vertical angle
measurement accuracy for machine engineering and geodetic instruments is
based mainly on the use of multiangle prism--polygon [4, 7] or precision
rotary encoder connected to the axis of the test bench and equipped with
an autocollimator placed in front of the axis of optical tube of the
instrument. Such equipment is used at the most famous optical companies
producing geodetic instruments with vertical angle measuring facilities.
Some inconveniences here also exist, the first one being of very
expensive arrangement to be constructed and produced with expensive
reference measure used in it. Using the big number of autocollimators
also is not convenient, as they must be placed at different (fixed)
angular positions and must be supplied of coaxial attached rotary
encoder or polygon for readings to be taken. One of advantages of such
arrangement is that the instrument to be checked is pointed to the
reticule of the autocollimator, in such way ensuring the strict
direction of the instrument's optical axis pointed to the infinity
of the autocollimator axis.
[FIGURE 2 OMITTED]
2. Arrangement for vertical angle calibration
The proposed arrangement for vertical angle calibration is based on
the trigonometric angle determination using the reference scale of the
length for vertical readings by the tacheometer and another reference
measure of length--for distance from the tacheometer's axis to the
vertical scale determination. The arrangement for calibration is shown
in Fig. 2.
The designations in Fig. 2 are: [I.sup.1]--initial
instrument's position with the axis of rotation of the spyglass O1;
[I.sup.2]--auxiliary instrument's position with the axis of
rotation of the spyglass [O.sub.2] This position is achieved by moving
the instrument along the slideways of the test bench for geodetic
instruments testing [7, 8]. At the distance l from the axis of the
instrument the linear scale S is fixed in vertical position to the
instrument's horizontal axis. The distance from the
instrument's both positions [l.sub.e] is fixed by using reference
measure of length, for example, end length gauge (length standard). The
linear photoelectrical transducer, laser interferometer or even precise
linear optical scale (with microscope) can be used for this purpose. It
is used for the determination of the distance from the axis of the
instrument to the surface of the scale, as it is quite complicated task
to do that initially. For the precise vertical angle measurements the
distance from instrument to be calibrated (tacheometer) and the
reference measure (linear scale) [l.sub.m] has to be determined quite
precisely (down to 0.01 mm) [9-11]. The accuracy of distance
determination influences the results of measurements considerably giving
a bias of reference data [3].
At the position [I.sup.1] of the instrument the reading h'
from the scale is taken at the angle [phi] of the axis of telescope of
the instrument and horizontal line.
The reading hh from the scale is taken. The angle of interest is
expressed
[phi] = arctg h'/l (1)
After displacement of the instrument linearly to the subsidiary
positions [I.sup.2] keeping the same vertical angle p, the next reading
h" is taken and the distance l can be determined
[phi] = arctg (h"/l + [l.sub.e]) (2)
and substituting the equations (1) and (2) will yield to
[MATHEMATICAL EXPRESSION NOT REPRODUCIBLE IN ASCII.] (3)
After taking the readings h and h from the scale S, the true value
of the distance l will be determined. Further measurements can be
performed determining every tested vertical angle of the instrument
operating with known distance and using the readings [h.sub.i] from the
scale. A full range of vertical angles of the geodetic instrument can be
tested at laboratory environment in such way improving the accuracy of
calibration and with a possibility to perform this at every desired time
in spite of meteorological conditions.
3. The experiment
The test of calibration of vertical angle measurements of the
geodetic angle measuring instrument (tacheometer) was performed. The
calibrated instrument Trimble 5503 tacheometer with the stated standard
deviation of angular measurements of 5" (arc sec) was used.
The arrangement for the experiment was composed according to Fig.
2; the tacheometer was mounted on the linear slideways and aimed to the
linear scale positioned vertically at a distance of approximately 2.5 m
for the tacheometer. The industrial laboratory linear scale of high
accuracy f 1 m in length with the scale strokes at every 1 mm was used.
The linear displacement of the tacheometer to be calibrated was
performed using the end length gauge of 200 mm. Optical scale of 1 m in
length was used for other tests. After the measurement and calculation
of linear distance from tacheometer to the scale performed according to
the previous chapter, it was determined that the distance l equals to
2.4215 m.
The main objective of experiment was to test the calibration method
and obtain preliminary results of the systematic errors (biases) of the
vertical angle measurements using Trimble 5503 tacheometer.
The calibration was performed using 1 m precise linear scale
collimating the spyglass to the scale strokes at a pitch of 10 mm and 12
mm. The resulting calculated deviations of tacheometer readings are
shown if Fig. 3 (10 mm pitch) and Fig. 4 (12 mm pitch).
[FIGURE 3 OMITTED]
[FIGURE 4 OMITTED]
Evaluate of the standard deviation of a collimation error (pointing
to the scale stroke) is equal to 2.15", when the evaluate of
general standard deviation of entire measurement procedure is 2.31"
[9]. Experimental standard deviation [1] of a vertical angle amounts to
[MATHEMATICAL EXPRESSION NOT REPRODUCIBLE IN ASCII.] (4)
where [r.sub.i] is the sum of squares of the residuals of the i th
series of measurements; [v.sub.i] is the number of degrees of freedom,
which is determined as
[v.sub.i = (n-1)t
where n is sets of vertical angles; t is targets, in our case--the
strokes of the reference linear scale.
As can be seen from Fig. 3 and 4, there can hardly be the short
period systematic constituent noticed according to deviations (though
there should be more measurements performed to finally state that).
Still there can be noticed a tendency for decrease of the accuracy
towards the limits of the measurements (-10[degrees] and 12[degrees]),
which is quite common for the tacheometers both due to the errors of
collimation at steep angles and the principle of action of vertical
angle encoder of the tacheometer though more tests should be performed
at that item too.
4. Conclusions
1. A simple method of calibration of vertical angle measurements
accuracy of geodetic instruments was proposed.
2. Similar method of angle measurement calibration could be
implemented for all angle measurements, both vertical and horizontal of
collimated instruments such as theodolites, tacheometers, etc.
3. The calibration of accuracy of vertical angle measurement of
Trimble 5503 tacheometer was performed using 1 m precise linear scale
positioned at a distance of 2.4215 m to the instrument to be calibrated.
4. Results of the calibration show real biases of measurement which
give an information of the accuracy parameters of the instrument.
Acknowledgments
This work has been funded by the Lithuanian State Science and
Studies Foundation, Project B-32/2009.
Received July 24, 2009
Accepted September 30, 2009
References
[1.] ISO 17123-5:2005. Optics and optical instruments--Field
procedures for testing geodetic and surveying instruments--Part 5:
Electronic tacheometers. Geneva: ISO, 2005.-11p.
[2.] Walser, B.H. Development and Calibration of an Image Assisted
Total Station: Dissertation-Zurich: ETH, 2004. -190p.
[3.] Ingensand, H. A new method of theodolite calibration.-XIX
International Congress.-Helsinki, Finland, 1990, p.91-100.
[4.] Jablonski, R., Shimokohbe, A., Nagai, A. Calibration system
for precision angle standards.-Bull. P.M.E. (T.I.T.), No45, 1980,
p.17-24.
[5.] Giniotis, V. Position and Displacement Measurements. -Vilnius:
Technika, 2005.-215p. (in Lithuanian).
[6.] Cooper, M.A.R. Modern Theodolites and Levels. Second ed. BSP
Professional Books, 1987.-197p.
[7.] Giniotis, V., Rybokas, M., Petroskevicius, P. Investigations
into the Accuracy of Angle Calibration. -Geodesy and Cartography, 2004,
Nr.3(30), p.65-70.
[8.] Bansevicius, R., Giniotis, V., Knight, J. The application of
mechatronics for machines accuracy improvement. -Mechanika. -Kaunas:
Technologija, 2001, Nr.3(29), p.44-49.
[9.] Bru?as, D., Giniotis, V. Circular scale eccentricity analysis.
-Mechanika. -Kaunas: Technologija, 2008, Nr.5(73), p.48-53.
[10.] Brucas, D., Suziedelyte-Vysockiene, J. Limited accuracy
reference free angular position determination. -Mechanika. -Kaunas:
Technologija, 2009, Nr.1(75), p.67-72.
[11.] Marcinkevicius, A.H. Automatic control of longitudinal form
accuracy of a shaft at grinding. -Mechanika. -Kaunas: Technologija,
2008, Nr.1(69), p.63-68.
V. Giniotis *, D. Brucas **, L. Siaudinyte ***
* Vilnius Gediminas Technical University, Sauletekio al. 11, 10223
Vilnius, Lithuania, E-mail: gi@ap.vtu.lt
** Vilnius Gediminas Technical University, Sauletekio al. 11, 10223
Vilnius, Lithuania, E-mail: domka@ktv.lt
*** Vilnius Gediminas Technical University, Sauletekio al. 11,
10223 Vilnius, Lithuania, E-mail: lauryna@inbox.lt