The CVD gas systems for carbon nanotube growth.
Mantha, Srinivas ; Vathsal, S.
Introduction--CVD Oven
The oven used for CVD is a tube furnace from carbolite [1]. The
oven chassis provides the power supply for the cylindrical oven which
contains a 300 mm long ceramic work tube with an inner diameter of 38
mm. The oven is heated with a resistance wire wound around the work
tube. The heated length is 250 mm long. The maximal reachable
temperature is 1200[degrees]C. A quartz tube with a length of 1 m and a
diameter of 30 mm is placed within the boring of the work tube and
closed on both sides with stainless steel plates (figure 1). Ring clips
are mounted on the tube on both sides and three screws with wing nuts
are used to fix the plates. This mounting is very delicate, since the
tube breaks when the ring clips are tightened too much, and the system
leaks when they are not tightened enough. A heat shield inside of the
quartz tube reduces the heat flow out of the oven.
[FIGURE 1 OMITTED]
The uniformity of temperature within the tube is tested. When the
tube is closed on both sides then the temperature fluctuates by [+ or -]
10[degrees]C expect for the last 10 cm at both ends. The fluctuation
within the middle zone of 95 mm is only [+ or -] 5[degrees]C. The
samples are placed in the middle of the oven on a quartz sample holder
with a length of 70 mm. This means that it might be assumed that the
temperature is approximately equal for all samples grown in one run.
The inlet of the oven is connected to the gas system with a
flexible metal tubing. The outlet of the oven and a water cooler are
joined with a short piece of teflon tube, since a plastic tube would be
destroyed by the hot and aggressive exhaust gas. The cooler is connected
with a piece of plastic tube to a bubbler and this again to the outside.
The cooler is used to reduce the temperature of the exhaust gas to
prevent the damage of the plastic tube. It is only necessary when
methane with a flow of 5 lit/min is used, and when the growth
temperature is higher than 1050[degrees]C. The cooler is not used for
lower growth temperatures and flows. The presence of bubbles within the
bubbler indicates that there is no major leak in the system. However
there is a more important effect of the bubbler, since it serves as a
diffusion barrier. There is always some unwanted gas diffusion into the
oven since the system is open to the outside. This flow is reduced since
it has to pass the water inside the bubbler. This means there is surely
some water vapour inside the reaction tube however the amount is limited
to the vapour pressure of water. Other gases can pass the bubbler but
they have to overcome two water gas boundary layers with areas of only 1
[cm.sup.2] and 5 [cm.sup.2], the surface of the bubble and the water air
boundary of the bubbler, respectively. The reaction tube is flushed with
argon during heating (around 30 minutes) which reduces the content of
environmental gases which are present due to the fact that the tube has
to be opened to mount the sample. However there is always a small
quantity of water vapour, nitrogen, oxygen and the other trace gases of
the environment present in the oven during growth in addition to the
used reaction gases. The reaction tube is flushed with argon (and in the
actually used process with hydrogen too) during cooling-down to prevent
the oxidation of the grown nanotubes [2].
The Gas Systems
The gas system provides the oven with the desired reaction gases.
Each of the presented gas systems contains at least one flowmeter which
allows the correct setting of the desired gas flow. All gas systems
consist of several gas lines for different gases which join before the
oven. The origin of each gas line is a gas bottle equipped with a
reduction valve which allows the setting of the desired gas pressure.
The next part is a one-way valve which protects the connected gas bottle
in the case of an explosion within the gas line. It opens only when the
pressure from the side of the gas bottle is higher then the pressure
from the other side. The pressure stroke of an explosion would close the
one-way valve and prevent the hot explosion gases from penetrating the
gas bottle. A following two way valve permits the fast opening and
closing of the gas line without changing the settings of the dedicated
reduction valve and flowmeter. Manometers before and after the flowmeter
allow the setting of a desired pressure drop over the flowmeter. The
different parts are connected with stiff stainless steel tubes with an
outer diameter of 6 mm. With one exception: more flexible tubes with an
outer diameter of 3 mm are used to connect the reduction valves with the
one way valves to guarantee the necessary flexibility which is needed
when the gas bottles have to be changed [2].
Gas System I, with One Variable Area Flowmeter
This is the most simple setup which is used (Figure 2.1). The gas
lines for all used gases (argon, methane and ethylene) are connected to
the same variable area flowmeter. The gas flow is controlled by closing
and opening the valves. The gas lines have an overall length from the
gas bottles to the oven of approximately 3 m [3,4].
Used gases: argon, methane and ethylene.
[FIGURE 2.1 OMITTED]
Gas System II, with Three Thermal Profile Flowmeters and a Variable
Area Flowmeter
This setup (Figure 2.2) belongs to the CVD oven mentioned above. A
Brooks instrument controller controls three Brooks instrument thermal
profile flowmeters gauged to Ar, H2 and N2. An additional variable area
flowmeter can be used for high flows of methane. The gas system is
connected to the oven with a 12 m long tube of a diameter of 3 mm. The
gas lines have an overall length from the gas bottles to the oven of
approximately 15 m [4,8].
Used gases: argon, hydrogen, methane, ethylene and acetylene.
[FIGURE 2.2 OMITTED]
Gas System III, with Two Variable Area Flowmeters, a Needle Valve
and a Switch
The distinctiveness of this setup (Figure 2.3) is a switch which is
used to toggle between argon and the carbon feedstock, which is usually
methane. For some experiments a small amount of ethylene is added to the
methane flow with a needle valve. The argon and the methane gas lines
were equipped with variable area flowmeters. The gas flow is controlled
by using the switch. It has two inlets (for argon and methane) and two
outlets (to the oven and outward). The switch has two positions. In
position 1 the methane line is connected to the oven and the argon line
is connected outward. In position 2 the argon line is connected to the
oven and the methane line is connected outward. The pressure after the
switch is [approximately equal to]0.1bar [+ or -] 10% relative (due to
the fluctuations of the air pressure from day to day) and the pressure
before the flowmeters is set to 0.2 bar relative using the manometers
(this corresponds to a pressure drop of [approximately equal to]0.1 bar
over the flowmeters). A special feature of this system is a needle valve
which is used to add a small quota of ethylene to the methane flow. The
gas lines have an overall length from the gas bottles to the oven of
approximately 3 m [7,8].
Used gases: argon, methane and ethylene.
[FIGURE 2.3 OMITTED]
Gas System IV, with Three Variable Area Flowmeters
This is the currently used gas system (Figure 2.4). It allows the
simultaneous use of argon, methane and hydrogen. All gas lines are
equipped with variable area flowmeters. Manometers before and after the
flowmeters allow the exact setting of the pressure drop over the
flowmeters (the pressure after the flowmeter depends from the actual air
pressure and fluctuates by maximally 10% from day to day). Since the
flowmeters are gauged to 0.2 bar the pressure drop over the flowmeters
is set to 0.2 bar, using the reduction valves of the gas bottles
considering the readout of the manometers in front of each flowmeter.
This makes a pressure correction unnecessary. A correction for the used
gas is necessary since the flowmeters are gauged to air (A correction
table is usually used). The gas lines have an overall length from the
gas bottles to the oven of approximately 3 m [4]. Used gases: argon,
hydrogen and methane.
[FIGURE 2.4 OMITTED]
Conclusion
The Gas System provides the oven with the desired reaction gases.
Gas System IV has three variable flowmeters for argon, hydrogen and
methane. The first experiment was performed by Gas System I which proved
the possibility to grow Carbon Nanotubes with the CVD oven. However, it
does not allow the simultaneous use of two or more gases. The Gas System
II belongs to the old CVD system and was connected to the oven with a
12m long tube of a diameter of 3 mm. Gas System III is less
sophisticated than Gas System II and allows flow of only two gases
methane and argon. The Gas System IV is currently in use now-a-days.
References
[1] Manual. MTF Tube furnaces: Installation, Operation and
Maintenance Instructions. (MTF--Carbolite.pdf on CD).
[2] M. Altendorf. Flow Handbook. Endress-Hauser, 2004.
[3] Manual: Low Volume Flowmetes and Switches by Kobold, KOBOLD
Messring GmbH, Nordring; url: www.koboldmessring.com
[4] Manual: Variable Area Flowmeter and Switches by Kobold, KOBOLD
Messring GmbH, Nordring; url: www.koboldmessring.com
[5] Manual: Paddle Type Flowmeter and Switches by Kobold, KOBOLD
Messring GmbH, Nordring; url: www.koboldmessring.com
[6] Manual: Rotating Vane Flowmeter and Switches by Kobold, KOBOLD
Messring GmbH, Nordring; url: www.koboldmessring.com
[7] Manual. Control and Read Out Unit, Models 0152/0154. Brooks,
1999.
[8] Manual. Model 5850E Mass Flow Controller. Brooks
(5850-E-BROOKS.pdf on CD), 1997.
[9] C. Schonenberger and L. Forro, "Physics of Multiwall
Carbon Nanotubes" Physics World Vol 13, No 6, 37-41 (2000)
[10] Edited by M.S. Dresselhaus, G. Dresselhaus and Ph. Avouris,
Carbon Nanotubes, Springer, Berlin Heidelberg, New York 2001
[11] L. Forro and C. Schonenberger, "Carbon Nanotubes,
Materials for the Future" Europhysics News 32, No. 3 (2001)
Srinivas Mantha, FIE, FIETE
Professor and HOD, Department of Electrical & Electronics
Engineering,
Vignan Institute of Technology and Science, Deshmukhi, Nalgonda
Dist.,
(Hyderabad). 508284 A.P., India.
E-mail: srin.mantha@gmail.com
S. Vathsal, FAeSI, FIETE
Retd. Scientist 'G', Director ER & IPR, DRDO,
Dean, School of Science and Humanities,
VIT University, Vellore.--632014 T.N., India.
E-mail: svathsal@gmail.com