Polynomial error approximation of a precision angle measuring system.
Siaudinyte, Lauryna ; Brucas, Domantas ; Rybokas, Mindaugas 等
Introduction
Precision angle measuring systems are widely used for manufacturing
various types of machinery embedding them into devices and instruments
such as total stations, metal cutting machines, etc. An angle encoder is
the main part of an angle measuring system. They may vary in sizes,
types and accuracy. Depending on the required accuracy, the whole angle
measuring system must be accordingly calibrated. Calibration of such
systems is essential to determine errors and increase measurement
accuracy.
Calibration process is frequently a very complicated procedure
requiring specific equipment, including standards for achieving the best
results. Precision angle measurement is compulsory for engineering at
various levels of precision, and therefore many artefacts and
instruments such as angle gauge blocks, optical polygons and rotary
tables are used for angle metrology. To ensure high reliability, they
should be calibrated using standard instruments such as an indexing
table along with an autocollimator (Just et al. 2003; Kim et al. 2011).
1. Equipment and method used for flat angle calibration
To determine the accuracy of an angle measuring system, a
comparative method was chosen. This method is based on a comparison of
the measurand and standard angle determining a deviation between those
angles. Equipment selected for realisation of this method consists of a
precision angle comparator, 36 mirror-sided polygon and an
autocollimator. The angle standard as a mirror polygon is an important
measurement object used in applied for an angle calibration world wide.
When measuring the angle accuracy of a product, many application fields
of using the polygon could be used (Watanabe et al. 2003). The angle
comparator consists of the basis that includes a precision
mechatronicrotary device, a device for detecting circular scale
graduation and its relative rotation according to the circular scale
measuring system. A section view of the angle comparator is presented in
Fig. 1. Another critical element is a calculation system for both the
control of the calibration process and data processing. The basic
mechatronic system is designed for direct limb and angular encoder
calibration. The basis of such system consists of a massive small-grain
grey granite brick with aerostatic rotational mechanism mounted on it
through a suspension ring. It is rotated using a worm gear. To ensure
proper rotation of this gear, combined aerostatic and rolling bearings
are installed. The worm gear is mounted on aerostatic bearings and
supported radially by rolling bearings. The rotation axis approximately
matches the axis of the aerostatic rotational device. For this reason,
suspension and radial bearings are mounted on position-correction
devices. The whole system is covered with a special cover and a
precisely finished top surface protecting the system of damage coming
from environmental pollutants. Additional measurement equipment can also
be placed on this surface (Kasparaitis, Sukys 2008).
[FIGURE 1 OMITTED]
There are two diverse requirements for the gear of the comparator:
it has to ensure constant angular rotation and the possibility of
precise positioning while providing a stable angular position within
given time.
The gear has not to create any other forces except for torque or
tangential rotational force. It is also important that the gear does not
generate intensive thermal activity or vibration. To perform accurate
measurements, the laboratory should be isolated from any external
vibrations under stable temperature.
Important parts of an angle measuring system are shown in the
section view in Fig. 1: 1-precision aerostatic shat, 2-reading head,
3-reference circular scale, 4-measured table, 5-rolling bearing, 6-worm
gear.
One of the variants of the gear that meets the requirements
mentioned before is a worm gear offering autonomous mechanism for the
rotation of the gear combined with a precise connection with a rotation
device that is rigid along the tangent direction of the spindle and
slender in other directions.
The examined comparator has an elastic worm gear connection with
the spindle generating pure torque without any radial or axial forces
and ensures the stability of the spindle. The worm gear is rotated by an
electric motor through a mechanical reducer and can be disconnected to
rotate the spindle manually.
The limb of the angle measuring system has a rigid connection
avoiding any intermediate sliding mechanical parts. It helps in
eliminating negative effects of hysteresis and reverse errors that have
a negative effect on precision and are common to mechanical coupling
devices (Kasparaitis, Sukys 2008; Sydenham, Thorn 1992).
There are two precise guiding devices made of small-grain granite
mounted on the basis of the comparator that have two sliding carriages
with stroke-detection microscopes mounted on them. They are placed
orthogonally to the axis of the spindle on aerostatic supports.
A 36 sided polygon is placed on the top of the precision angle
measuring system with the embedded Renishaw angle encoder. The system is
precisely centred and levelled. The autocollimator Hilger & Watts is
pointed directly to the mirror face of the polygon and the system is set
to the position in which the readout of the autocollimator is 0.00 arc
seconds as shown in Fig. 2.
[FIGURE 2 OMITTED]
2. Data processing
A systematic error along with the evaluation of its possible cause
is one of the key procedures in data processing (Rabinovich 2010). One
of the main features influencing the precision of any rotary device is
the eccentricity of mounting a rotating or measuring element
(eccentricity of bearing or the measuring scale etc.). Having obtained
scale calibration results, the eccentricity of the scale (or a disc
mounting) can be calculated. Data on encoder systematic errors allow
determining the mechanical systematic errors of the elements of the test
rig such as the eccentricity of bearings or the scale itself (Kim et al.
2011, Stone et al. 2004). The encoder was calibrated using one reading
head and the first harmonic was noticed in the chart. The results of
calibrating the encoder are displayed in Fig. 3.
[FIGURE 3 OMITTED]
[FIGURE 4 OMITTED]
The measurement results are displayed in Fig. 4. The measurement
errors were calculated following the evaluation of the results of
polygon calibration and the accuracy of the autocollimator. Two reading
sensors were embedded in the measuring system to eliminate eccentricity
errors. Fig. 4 displays the results of twelve circles rotating the angle
measuring system forwards and backwards.
Having the smoothened data and using Fourier analysis, the
harmonics of measurements could be calculated (Brucas, Giniotis 2008;
Brucas et al. 2006). Finite Fourier series can be expressed applying the
following formula:
[delta][phi]([phi]) = [A.sub.0] + 2 [n-1.summation over (m=1)]
([A.sub.m] cos 2[pi] m [f.sub.1][phi] + [B.sub.m] sin 2[phi] m
[f.sub.1][phi]) + [A.sub.n] cos 2[phi] m [f.sub.1][phi]. (1)
The coefficients of the equation for each type of harmonics can be
calculated as
[A.sub.m] = 1 / N [n-1.summation over (r=-n)] [delta][[phi].sub.r]
cos 2[pi]mr / N; (2)
[B.sub.m] = 1 / N [n-1.summation over (r=-n)] [delta][[phi].sub.r]
cos 2[pi]mr / N; (3)
where: m--the number of harmonics, N--the total number of
measurements, r--the measurement number.
The amplitude of each type of harmonics is
[R.sub.m] = [square root of ([A.sup.2.sub.m] + [A.sup.2.sub.m])].
(4)
The phase shift of each type of harmonics (regarding the zero
point) is
[[phi].sub.m] = arctg (-[B.sub.m] / [A.sub.m]) (5)
The received results indicate that the polynomial curve of the
fourth order reflects the view of the present measurement data. The
equation for the fourth order polynomial curve is
y = (7 x [10.sup.-4])[x.sup.4]--(4 x [10.sup.-7])[x.sup.3] + (6 x
[10.sup.-5])[x.sup.2] + 0,0022x--0,1824. (6)
The calculation of a standard deviation is shown in Fig. 5.
[FIGURE 5 OMITTED]
The equation for the polynomial curve of the standard deviation can
be expressed as
y = (6 x [10.sup.-11])[x.sup.4]--(2 x
[10.sup.-8])[x.sup.3]--[10.sup.-6][x.sup.2] + 0,0011x + 0,1003. (7)
Conclusions
Systematic errors of the angle measuring system have been
determined. Calculating the errors of the polygon and the accuracy of
the evaluated autocollimator and encoder has displayed the results of
systematic errors that tend to have the highest values in the range of
the 36 sided polygon of 100[degrees]-170[degrees]. This leads to the
assumption that obvious systematic errors may be caused by other
components of the angle measuring system (such as bearings) or the
accuracy of installing the measurement system.
doi:10.3846/20296991.2013.786879
Acknowledgement
This research was funded by the European Social Fund under the
Global Grant measure.
References
Brucas, D.; Giniotis, V. 2008. Circular scale eccentricity
analysis, Mechanika 2(57): 48-53. ISSN 1392-1207.
Brucas, D.; Giniotis, V.; Petroskevicius, P. 2006. The construction
of the test bench for calibration of geodetic instruments. Geodesy and
Cartography 32(3): 66-70. ISSN 13921541.
Just, A.; Krause, M.; Probst, R.; Wittekopf, R. 2003. Calibration
of high-resolution electronic autocollimators against an angle
comparator, Metrologia 40: 288-294.
http://dx.doi.org/10.1088/0026-1394/40/5/011
Kasparaitis, A.; Sukys, A. 2008. Metrologinio ukio ir matavimu
organizavimas. Vilnius: Technika. 192 p. ISBN 978-9-95528306-5.
http://dx.doi.org/10.3846/991-S
Kim, J. A., et al. 2011. Precision angle comparator using
self-calibration of scale errors based on the equal-division-averaged
method, in Proc. of MacroScale 2011. Wabern, 1-4.
Rabinovich, S. G. 2010. Evaluating measurement accuracy. New York:
Springer. 278 p. ISBN 978-1-4419-1455-2.
http://dx.doi.org/10.1007/978-1-4419-1456-9
Stone, J. A.; Amer, M.; Faust, B.; Zimmerman, J. 2004.
Uncertainties in small-angle measurement systems used to calibrate angle
artifacts, Journal of Research of the National Institute of Standards
and Technology 109(3): 319-333. http://dx.doi.org/10.6028/jres.109.024
Sydenham, P. H.; Thorn, R. 1992. Handbook of Measurement science:
Elements of Change, vol. 3. Chichester. New York. Brisbane. Toronto.
Singapore: John Wiley & Sons. 600 p.
Watanabe, T.; Fujimoto, H.; Nakayama, K.; Kajitani, M.; Masuda, T.
2003. Calibration of polygon mirror by the rotary encoder calibration
system, in Proc. of XVIIIMEKO World Congress, June 22-27, 2003,
Dubrovnik, Croatia, 18901893.
Lauryna Siaudinyte (1), Domantas Brucas (2), Mindaugas Rybokas (3),
Genadijus Kulvietis (4)
(1,2) Institute of Geodesy, Vilnius Gediminas Technical University,
Sauletekio al. 11, LT-10223 Vilnius, Lithuania
(3,4) Department of Information Technologies, Vilnius Gediminas
Technical University, Sauletekio al. 11, LT-10223 Vilnius, Lithuania
E-mails: (1) lauryna.siaudinyte@vgtu.lt (corresponding author); (2)
domka@ktv.lt; (3) mindaugas.rybokas@vgtu.lt; 4gk@vgtu.lt
Received 05 February 2013; accepted 26 February 2013
Lauryna SIAUDINYTE. Junior research fellow at the Institute of
Geodesy. Sauletekio al. 11, LT-10223 Vilnius, Lithuania, Ph. +370 5 274
4705, e-mail: lauryna.siaudinyte@vgtu.lt
MSc from VGTU in 2010. Currently, PhD student at the Department of
Geodesy and Cadastre, Vilnius Gediminas Technical University.
Research interests: calibration of geodetic instruments, angle
measurements, measurement accuracy and precision.
Domantas BRUCAS. Head of Space Science and Technology Institute,
Sauletekio al. 15, LT-10223 Vilnius, Lithuania. Doctor (2008). The
author of two educational books and more than 30 scientific papers.
Participated in a number of international conferences, owns three
registered patents.
Research interests: comparator development for angular
measurements, automation of processing measurement results.
Mindaugas RYBOKAS. Assoc. Prof., Dr at the Department of
Information Technologies, Vilnius Gediminas Technical University,
Sauletekio al. 11, LT-10223 Vilnius, Lithuania (Ph. +370 5 274 4832),
e-mail: mindaugas.rybokas@vgtu.lt.
The author of more than 35 scientific papers; participated in a
number of International conferences.
Research interests: analysis of information measuring systems.
Genadijus KULVIETIS. Prof., Dr Habil at the Department of
Information Technologies, Vilnius Gediminas Technical University,
Sauletekio al. 11, LT-10223 Vilnius, Lithuania. E-mail: gk@vgtu.lt.
The author of two educational books and more than 50 scientific
papers. Participated in a number of international conferences.
Research interests: analysis of information measuring systems.