首页    期刊浏览 2024年11月24日 星期日
登录注册

文章基本信息

  • 标题:Effect of Ambient Temperature on Calibration of Cooled Thermal Camera
  • 本地全文:下载
  • 作者:Amit Kumar Sharma ; Sanjay Kumar Sharma ; Prashant Vasistha
  • 期刊名称:Defence Science Journal
  • 印刷版ISSN:0976-464X
  • 出版年度:2017
  • 卷号:67
  • 期号:2
  • 页码:173-176
  • 语种:English
  • 出版社:Defence Scientific Information & Documentation Centre
  • 摘要:Thermal cameras may be used under ambient conditions that differ significantly from the calibration conditions. The effect of ambient temperature on temperature measurement error is examined for MWIR and LWIR cooled thermal cameras. The facilities used include an environmental chamber and an extended area blackbody with temperature controller. Significant differences were observed in the temperatures measured by the cameras placed in different ambient temperatures, with reference to the set blackbody temperatures. Re-calibration was done to account for variations in ambient temperature from 5 ºC to on the outputs of the cameras. It was found that after such recalibration, the measurement error was within acceptable accuracy of ±1 °C.
  • 关键词:Calibration;Calibration drift;thermal camera;Ambient temperature