文章基本信息

  • 标题:Repeatability and Validity of the VAA-2000 Automated Self Vision Tester
  • 本地全文:下载
  • 作者:Kee, Hyun Ju ; Yang, Hee Kyung
  • 期刊名称:Journal of the Korean Ophthalmological Society
  • 印刷版ISSN:0378-6471
  • 出版年度:2017
  • 卷号:58
  • 期号:2
  • 页码:203-207
  • DOI:10.3341/jkos.2017.58.2.203
  • 语种:Korean
  • 出版社:The Korean Ophthalmological Society
  • 摘要:Purpose

    To evaluate the test-retest reliability and validity of the VAA-2000 Self Vision Tester.

    Methods

    A total of 108 eyes of 108 adults (55 men and 53 women) who visited Seoul National University Bundang Hospital from March to December 2015 and who had best corrected visual acuities of 20/200 or better, according to both Hahn's visual acuity test (using numbers and Landolt rings) and the VAA-2000 Self Vision Tester, were included. The test-retest reliability and validity of the VAA-2000 Self Vision Tester were assessed and compared with Hahn's visual acuity test performed using numbers and Landolt rings. The intraclass correlation, Pearson correlation, Bland-Altman plots, and Deming regression between the different tests were analyzed.

    Results

    The average best corrected visual acuity was 0.12 ± 0.17 logMAR measured with the VAA-2000 Self Vision Tester, and was 0.08 ± 0.20 logMAR with Hahn's visual acuity test using numbers and 0.08 ± 0.30 logMAR using Landolt rings. The VAA-2000 Self Vision Tester demonstrated high test-retest reliability (intraclass correlation coefficient = 0.903, p < 0.001) and acceptable validity when compared with Hahn's visual acuity test using numbers (r = 0.767, p < 0.001) and Landolt rings (r= 0.727, p < 0.001).

    Conclusions

    The VAA-2000 Self Vision Tester possesses good reliability and acceptable validity when compared to Hahn's visual acuity test using numbers and Landolt rings.