  • 标题:Comparison of Intraocular Pressure via Goldmann-applanation Tonometry and TonoPen in Thyroid-associated Ophthalmopathy Accompanying Restrictive Strabismus
  • 本地全文:下载
  • 作者:Kim, Jun Sik ; Lee, Eun Ji ; Kim, Namju
  • 期刊名称:Journal of the Korean Ophthalmological Society
  • 印刷版ISSN:0378-6471
  • 出版年度:2017
  • 卷号:58
  • 期号:6
  • 页码:685-691
  • DOI:10.3341/jkos.2017.58.6.685
  • 语种:Korean
  • 出版社:The Korean Ophthalmological Society
  • 摘要:Purpose

    To compare the intraocular pressure (IOP) measured using Goldmann-applanation tonometry (GAT) and Tonopen® tonometry and to evaluate the factors influencing the measurement difference in patients with thyroid-associated ophthalmopathy (TAO)-related restrictive strabismus.

    Methods

    In 50 eyes of 50 patients who were diagnosed with TAO, IOP measurements were taken using both GAT and a Tonopen® and were subsequently compared between the devices. Factors influencing the measurement difference between the devices were determined, including the restriction of eyeball movement, eyeball deviation, exophthalmometry, central corneal thickness, refractive errors, and blood thyroid hormone levels.

    Results

    In the TAO patients, the GAT-measured IOP was higher than for Tonopen® (16.1 ± 4.7 vs. 13.8 ± 4.5 mmHg, respectively, p < 0.001). As the restriction of vertical eyeball movement increased, the IOP difference between the devices also increased ( p = 0.037). The absolute IOP difference between the devices was positively correlated with restrictions in vertical eyeball movement ( p = 0.027), degree of vertical strabismus ( p = 0.021), and central corneal thickness ( p ≤ 0.031).

    Conclusions

    In patients with TAO accompanying vertical eyeball movement restriction, potential errors in IOP measurements should be considered between the different IOP-measuring devices.

  • 关键词:Goldmann-applanation tonometry; Intraocular pressure; Restrictive strabismus; thyroid-associated ophthalmopathy; TonoPen®