期刊名称:Proceedings of the National Academy of Sciences
印刷版ISSN:0027-8424
电子版ISSN:1091-6490
出版年度:1974
卷号:71
期号:4
页码:1294-1298
DOI:10.1073/pnas.71.4.1294
语种:English
出版社:The National Academy of Sciences of the United States of America
摘要:A combination of deformable flow probes of negligible lateral dimensions with an electronic circuit capable of providing a prolonged plateau of dB/dt = 0 and of sampling the flow signal at the end of this interval permits electromagnetic measurement of blood flow with a reliable zero base line secured by switching off the magnet. An extracorporeal magnet provides the magnetic field. The flow transducer is introduced into the vascular system percutaneously through a standard angiographic catheter by conventional technique. The idea of the current generator can be described as "principle of interrupted resonance." The current wave form can be described as a sequence of disconnected bisected sine waves joined at the apices by horizontal current plateaus where di/dt is strictly zero.