摘要:We optimized proton-induced X-ray emission (PIXE) for tissue analysis in a toxicity-disposition study. We used cultured rabbit renal slices as the biological system to demonstrate the use of PIXE analysis. The renal slices were exposed to HgCl2, CdCl2, K2Cr2O7, or NaAsO2 alone or in a mixture. The PIXE analysis provides information on concentrations of elements above atomic number 11, and it is the only analytical technique that can determine 20-30 elements nondestructively in a single, small sample (approximately 5 mg) with detection limits of 1-5 ppm (dry weight). The renal slices are thin targets that yield X-ray emission spectra with low backgrounds and high elemental sensitivities. The nondestructive nature of PIXE and the ability to simultaneously measure uptake of multiple metals and endogenous elements are unique to this methodology. Full text Full text is available as a scanned copy of the original print version. Get a printable copy (PDF file) of the complete article (4.0M), or click on a page image below to browse page by page. Links to PubMed are also available for Selected References . 302 303 304 305 306 307 308