首页    期刊浏览 2024年10月06日 星期日
登录注册

文章基本信息

  • 标题:Influence of Leakage from Non-Invasive Positive Pressure Ventilation Mask on FiO2 Value Delivered by Home Oxygen Therapy Concentrator: A Bench Study on Simulating Patients with Chronic Obstructive Pulmonary Disease
  • 本地全文:下载
  • 作者:Kazuto Doi ; Mitsuhiro Nishitani ; Mayumi Doi
  • 期刊名称:Health
  • 印刷版ISSN:1949-4998
  • 电子版ISSN:1949-5005
  • 出版年度:2018
  • 卷号:10
  • 期号:07
  • 页码:919-927
  • DOI:10.4236/health.2018.107068
  • 语种:English
  • 出版社:Scientific Research Publishing
  • 摘要:Introductio n : During the application of non-invasive positive pressure ventilation (NPPV) therapy in home mechanical ventilation (HMV), leaks in the NPPV mask may occur owing to the position of the mask due to conditions such as skin disorders. Methods : To investigate whether such a leak affects FiO2 supplied to the alveoli, we simulated a patient with chronic obstructive pulmonary disorder during NPPV in HMV. In addition, FiO2 was measured in the portion assumed to be the mouth and lungs while setting the flow of the oxygen concentrator and leak amount based on a previous study. Results : FiO2 supplied to the lungs increased statistically significantly upon increasing the amount of leak (P Conclusions: We observed that FiO2 supplied to alveoli can be reduced by a leak in the NPPV mask. Because our results differ from those previously reported, we believe that further studies should reassess the selection of respirators and oxygen concentrators.
  • 关键词:Mask;Oxygen;Ventilation;Positive-Pressure Ventilation;Chronic Obstructive Pulmonary Disorder
国家哲学社会科学文献中心版权所有