摘要:In mass vaccination programmes, the jet-injection of vaccine may have considerable operational advantages over the classical techniques. The technical performance of two models of jet-injector, the Dermo-Jet and the Ped-O-Jet, in BCG vaccination was assessed in a number of studies which are reviewed by the authors. It is shown that the jet-injectors do not administer the full dose for which they are calibrated and that the size of the vaccination lesion varies more than after vaccination by syringe. By increasing the dosage considerably, the results of vaccination by jet-injection may be improved to a certain extent but the risk of unpleasant reactions is also increased. Full text Full text is available as a scanned copy of the original print version. Get a printable copy (PDF file) of the complete article (2.0M), or click on a page image below to browse page by page. Links to PubMed are also available for Selected References . 707 708 709 710 711 712 713 714 715 716 717 718 719 720