首页    期刊浏览 2025年06月19日 星期四
登录注册

文章基本信息

  • 标题:Performance of the Reflotron in Massachusetts' Model System for Blood Cholesterol Screening Program.
  • 本地全文:下载
  • 作者:S Havas ; R Bishop ; L Koumjian
  • 期刊名称:American journal of public health
  • 印刷版ISSN:0090-0036
  • 出版年度:1992
  • 卷号:82
  • 期号:3
  • 页码:458-461
  • 出版社:American Public Health Association
  • 摘要:The precision, accuracy, and durability of the Reflotron were evaluated by the Massachusetts Model Systems for Blood Cholesterol Screening Program. Screenings were conducted in diverse community settings over 16 months. Fingerstick samples from 10,428 individuals were tested. None of the four analyzers met the 1992 standards for precision, although two met the 1992 standards for accuracy. More than 40% of Reflotron values differed from the reference laboratory values by upwards of 5%. More than 16% of individuals were misclassified in terms of their risk category. All four instruments malfunctioned during the project.
国家哲学社会科学文献中心版权所有