首页    期刊浏览 2024年11月28日 星期四
登录注册

文章基本信息

  • 标题:Precision and accuracy of a portable blood analyzer system during cholesterol screening.
  • 本地全文:下载
  • 作者:P Greenland ; N L Bowley ; C A French
  • 期刊名称:American journal of public health
  • 印刷版ISSN:0090-0036
  • 出版年度:1990
  • 卷号:80
  • 期号:2
  • 页码:181-184
  • 出版社:American Public Health Association
  • 摘要:The precision and accuracy of two Kodak Ektachem DT-60 portable blood analyzers were assessed in a model (research) cholesterol screening program in Rochester, New York. Between June and October 1987, a total of 8,573 people underwent a cholesterol screening held in a movable trailer. A wide variety of temperature, humidity, and other potentially adverse conditions were encountered during the screening period. Between-run coefficients of variation ranged from 1.9 percent to 4.8 percent per month; average bias compared to a Reference Laboratory method ranged between 0.2 percent and 2.0 percent. Both precision and accuracy met currently recommended standards for cholesterol testing in the United States.
国家哲学社会科学文献中心版权所有