  • 标题:Spiral scanning imaging and quantitative calculation of the 3-dimensional screw-shaped bone-implant interface on micro-computed tomography
  • 作者:Choi, Jung-Yoo Chesaria ; Choi, Cham Albert ; Yeo, In-Sung Luke
  • 期刊名称:The Journal of the Korean Academy of Periodontology
  • 印刷版ISSN:0250-3352
  • 出版年度:2018
  • 卷号:48
  • 期号:4
  • 页码:202-212
  • DOI:10.5051/jpis.2018.48.4.202
  • 语种:English
  • 出版社:Korean Academy of Periodontology
  • 摘要:Purpose

    Bone-to-implant contact (BIC) is difficult to measure on micro-computed tomography (CT) because of artifacts that hinder accurate differentiation of the bone and implant. This study presents an advanced algorithm for measuring BIC in micro-CT acquisitions using a spiral scanning technique, with improved differentiation of bone and implant materials.

    Methods

    Five sandblasted, large-grit, acid-etched implants were used. Three implants were subjected to surface analysis, and 2 were inserted into a New Zealand white rabbit, with each tibia receiving 1 implant. The rabbit was sacrificed after 28 days. The en bloc specimens were subjected to spiral (SkyScan 1275, Bruker) and round (SkyScan 1172, SkyScan 1275) micro-CT scanning to evaluate differences in the images resulting from the different scanning techniques. The partial volume effect (PVE) was optimized as much as possible. BIC was measured with both round and spiral scanning on the SkyScan 1275, and the results were compared.

    Results

    Compared with the round micro-CT scanning, the spiral scanning showed much clearer images. In addition, the PVE was optimized, which allowed accurate BIC measurements to be made. Round scanning on the SkyScan 1275 resulted in higher BIC measurements than spiral scanning on the same machine; however, the higher measurements on round scanning were confirmed to be false, and were found to be the result of artifacts in the void, rather than bone.

    Conclusions

    The results of this study indicate that spiral scanning can reduce metal artifacts, thereby allowing clear differentiation of bone and implant. Moreover, the PVE, which is a factor that inevitably hinders accurate BIC measurements, was optimized through an advanced algorithm.

  • 关键词:Algorithm; Artifacts; Bone-implant interface; Partial volume effect; Spiral path; X-ray microtomography