首页    期刊浏览 2024年07月05日 星期五
登录注册

文章基本信息

  • 标题:Development and implementation of standardized method for detecting immunogenicity of acellular pertussis vaccines in Korea
  • 作者:Park, Chulmin ; Park, Chulmin ; Huhng Ho
  • 期刊名称:Clinical and Experimental Vaccine Research
  • 印刷版ISSN:2287-3651
  • 出版年度:2019
  • 卷号:8
  • 期号:1
  • 页码:35-42
  • DOI:10.7774/cevr.2019.8.1.35
  • 语种:English
  • 出版社:KoreaMed Synapse
  • 摘要:Purpose

    There is no standard method for confirming the immunogenicity of acellular pertussis vaccines. We tried to develop a local standard method for evaluating the immunogenicity of the three-component of acellular pertussis vaccines which was developed by a Korean local company.

    Materials and Methods

    The developed pertussis antigens (pertussis toxin, filamentous hemagglutinin, pertactin) were evaluated by in-house enzyme-linked immunosorbent assay (ELISA) using 189 negative sera, 25 positive sera, and 73 paired sera (pre- and post-Tdap [tetanus, diphtheria, and acellular pertussis] vaccinated sera). ELISA units were calculated by the reference line method, compared with World Health Organization reference sera, and the cut-off value was calculated using negative sera.

    Results

    When compared to National Institute for Biological Standards and Control control antigen (NIBSC) control antigens, the developed pertussis toxin (PT) and filamentous haemagglutinin (FHA) antigens were 203.48 and 61.60 IU/µg, respectively. Each in-house ELISA was established by validating the coefficients of variation % (PT, 11.53%; FHA, 8.60%; pertactin [PRN], 9.86%) obtained from the results of inter- and intra-assay variation. Also, the cut-off values of PT, FHA, and PRN were 11.65, 38.95, and 5.66 EU/mL, respectively. The distributions of antibody levels in paired showed that 93.15% (68/73) in anti-PT IgG, 97.26% (72/73) in anti-FHA IgG, and 100% in anti-PRN IgG were higher than a 100% increase after vaccination. Additionally, the values of 89.04% (65/73) in anti-PT IgG, 97.26% (72/73) in anti-FHA IgG, and 100% in anti-PRN IgG were below each cut-off point.

    Conclusion

    We established an in-house ELISA method using self-developed antigens, and these immunoassays have provided a way to standardize measuring the immunogenicity of newly developed vaccines, through single- and dual-serology.

  • 关键词:Pertussis Vaccine; Pertussis toxin; FHA protein; Pertactin; Enzyme-Linked Immunosorbent assay; Serology