首页    期刊浏览 2024年09月02日 星期一
登录注册

文章基本信息

  • 标题:An anti-clogging method for improving the performance and lifespan of blood plasma separation devices in real-time and continuous microfluidic systems
  • 本地全文:下载
  • 作者:Dong-Hyun Kang ; Kyongtae Kim ; Yong-Jun Kim
  • 期刊名称:Scientific Reports
  • 电子版ISSN:2045-2322
  • 出版年度:2018
  • 卷号:8
  • 期号:1
  • 页码:17015
  • DOI:10.1038/s41598-018-35235-4
  • 语种:English
  • 出版社:Springer Nature
  • 摘要:On-chip blood plasma separators using microfluidic channels are typically developed as disposable devices for short-term use only because blood cells tend to clog the microchannels, limiting their application in real-time and continuous systems. In this study, we propose an anti-clogging method. We applied dielectrophoresis to prevent microchannel clogging in a plasma separator that can be used over long periods for real-time and continuous monitoring. Prior to applying the anti-clogging method, the blood plasma separator stopped working after 4 h. In contrast, by manipulating the separator with the new anti-clogging method at a voltage of 20 V, it continued working in a long-term experiment for 12 h without performance deterioration or an increase in cell loss. Two critical performance parameters of the manipulated separator, the purity efficiency and the plasma yield, were 97.23 ± 5.43% and 38.95 ± 9.34%, respectively, at 20 V after 15 min. Interestingly, the two performance parameters did not decrease during the long-term experiment. Hence, the blood plasma separator with the anti-clogging method is an interesting device for use in real-time and continuous blood plasma separation systems because of its consistent performance and improved lifespan.
国家哲学社会科学文献中心版权所有