  • 标题:Suitability of various chromatographic and spectroscopic techniques for analysis and kinetic degradation study of trelagliptin
  • 本地全文:下载
  • 作者:Wafaa A. Zaghary ; Shereen Mowaka ; Mostafa A. Hassan
  • 期刊名称:Scientific Reports
  • 电子版ISSN:2045-2322
  • 出版年度:2017
  • 卷号:7
  • 期号:1
  • 页码:17255
  • DOI:10.1038/s41598-017-17642-1
  • 语种:English
  • 出版社:Springer Nature
  • 摘要:Multifaceted comparative analytical methods for trelagliptin (TRL) were investigated, applied to ZAFATEK tablets and HPLC-UV was selected for a degradation kinetic study. UPLC-MS/MS (Method I), UPLC-UV (Method II), HPLC-UV (Method III), UHPLC-UV (Method IV) and direct UV (Method V) methods were developed. Methods (I-V) showed satisfactory results using TRL concentration ranges of 50-800 ng/mL, 2.5-80 μg/mL, 5-100 μg/mL, 5-100 μg/mL and 5-50 μg/mL, respectively. Multiple Reaction Monitoring (MRM) of the transition pairs of m/z 358.176 to 134.127 for TRL and m/z 340.18 to 116.08 for alogliptin (IS) were employed utilizing positive mode Electrospray Ionization (ESI). The degradation kinetic study (Method VI) was carried out using 1 N HCl based on three different temperatures (70 °C, 80 °C and 90 °C). Through the optimized method-3, a good chromatographic separation of TRL from its major degradation product was achieved. Arrhenius plot was used in the kinetic study and the apparent 1st order degradation rate constant (K), t1/2, t90, and the activation energies were calculated for each temperature and at 25 °C. The optimized UPLC-MS/MS method is suitable for further TRL assay either in biological fluids or in the presence of impurities.