Vacuum Gauge Performance Verification System.
Stekleins, Antons ; Gerins, Eriks ; Kromanis, Artis 等
Vacuum Gauge Performance Verification System.
1. Introduction
One of the vacuum system main requirements is to achieve vacuum in
the system and maintain it permanently unchanged to a certain level
during the technological process, especially during a coating process.
Typically, there are two reasons for providing a vacuum: firstly, to
control the reactivity of the deposition process and, secondly, to
reduce the number of gas particles collisions thus enabling higher
energy particles to be used as part of the deposition process [2].
Vacuum gauge calibration is commonly used when pressure measurement
accuracy starts to deviate and it is impossible to secure technological
process stability. Some studies analyze vacuum gauge accuracy and
calibration period, but there are no clear approaches for vacuum gauge
verification under manufacturing conditions before calibration
operation. [1,5]. Common to all measuring devices, a pressure-measuring
instrumentation needs periodic calibration, primarily to monitor changes
in performance [1]. Reliable calibration of gauges is impossible without
establishing a widely recognized, true models of accuracy, to which
gauges can be verified, also known as "primary vacuum
standards" [5]. Therefore we are at the opinion that calibration
period must be determined individually by applying specific methodology
which includes multifunctional vacuum gauge verification using our
developed vacuum system.
In this paper we analyze vacuum gauge unreliable pressure
measurements during technological coating process at Sidrabe Inc. and
present developed vacuum system designed for vacuum gauge verification
under manufacturing conditions. The system provides fast, reliable and
multifunctional vacuum gauge verification. Ionization vacuum gauges were
tested by comparison method using custom made vacuum system.
Main purpose of the paper is to propose a new vacuum verification
system which can be used in any vacuum coating company to verify vacuum
gauges at any moment, even daily to obtain vacuum gauge pressure
deviation data.
2. Vacuum gauge pressure deviation analyses
At present, there have been identified two core problems associated
with the technological coating processes: firstly, a vacuum gauges
pressure measurement deviation and, secondly, frequency and the
necessity of the calibration.
During manufacturing calibrated vacuum gauges are being used to
ensure all necessary product characteristics and quality. In general,
the vacuum system contains minimum of two different types of vacuum
gauges, which are used to cover several ranges of vacuum, but
nevertheless a comparison between two gauges indicated pressure should
be performed. Prior art discusses a fact that vacuum gauge accuracy can
be influenced by several factors such as: operating conditions
(vibration, contamination, and temperature), connection, mounting and
corrosive gases [5, 10, 9]. Main problem we have encountered during a
daily coating process at Sidrabe Inc. is that the vacuum gauges
indicated pressure deviates after vacuum chamber evacuation. The
deviation is illustrated in Fig. 1.
Mainly two types of vacuum gauges were used: INFICON[R] diaphragm
vacuum gauge under number three at Fig. 1 (a, b) and ionization gauge
HPG 400 under number two at Fig. 1 (a,b) [8]. Indicated pressure
measurements by two vacuum gauges differ and it makes it even harder to
rely on any gauge with 100% confidence. As it is already known that the
diaphragm vacuum gauge is not affected by gas, therefore it is more
reliable than ionization gauge [4, 8]. In this experiment as illustrated
in Fig. 1 (a) an argon is fed into the chamber in volume of 100 sccm or
14,4%. [8]. In order to reach a pressure measurement without correction
and change it from 8,58 up to 6,88-4 torr, argon must be reduced to 11,2
% or 78 sccm [8].
If diaphragm vacuum gauge indicated pressure would be considered as
true value due to it's resistance to gas, then difference between a
true value and ionization gauge indicated pressure with correction
turned on would be 59,5%. Current experiment shows used vacuum gauge
indicated pressure deviations influence on argon flow amount needed for
the process in order to reach required pressure. One of the main
problems is that after some time vacuum gauges accuracy start to deviate
and it becomes even harder to determine which vacuum gauge indicated
correct measurements. Usage of unreliable and inaccurate vacuum gauge
may result in following consequences: poor product quality, loss of
material, resources and production time for repeating a process
(defects). Even minor deviations may leave a significant impact on
product characteristics and therefore it's crucial to have reliable
vacuum gauge with high accuracy and repeatability. If vacuum system has
one vacuum gauge connected, then a determination of gauge pressure
precision becomes even more problematic due to lack of comparison
operation when two vacuum gauges are connected.
Another issue is related to vacuum gauge calibration necessity and
frequency. Prior art discusses that this issue can be addressed with
certainty [1]. For majority of instruments the recalibration interval is
approximately a year, but if no historical data are available, which
relates to the device when used in a manufacturing environment and in a
way that is similar to the way it is normally used, then it is not
possible to answer the question [1]. Main problem is that during a daily
coating process it is possible to reach a point when product
characteristics fall outside the permissible limits due to vacuum gauge
accuracy failure. Since vacuum gauge operating time, environment inside
a vacuum system and other factors are up to the specific coating
process, then for the manufacturer it becomes even harder to specify
calibration period. If manufacturer suspects that vacuum gauge is
defective and unreliable, then the suspected gauge must be disconnected
from the vacuum system and sent to the laboratory where it will be
calibrated. Usually the manufacturer cannot afford to waste time and
resources for calibration operation due to manufacturing intensity,
therefore it's necessary to come up with the methodology of vacuum
gauge performance verification under production conditions. Using
developed methodology would give a possibility to verify a vacuum gauge
measurement accuracy, detect deviations in measurements and perform test
operations to ensure its accuracy. Since developed system allows all
necessary operations to be performed under manufacturing conditions it
will facilitate the decision- making about calibration necessity.
3. Custom made vacuum verification system
The aim of the experiment was to build a small size vacuum system
in order to reach high vacuum and perform all manipulations needed for
vacuum gauge verification using comparison method, gas flow, correction
factor and to obtain the necessary data. Custom made vacuum verification
system at Sidrabe Inc. can be seen in Fig. 2.
The vacuum system was design, manufactured and adjusted in two
years meeting all selected criteria for fast system evacuation with
steady flow of the gas and specific positioning of the vacuum gauges.
Vacuum system consists of following components: a chamber, a gate valve,
a mechanical pump and a turbo pump, solenoid valves, a manual leak
valve, a mass flow controller, flanges with connection ports for vacuum
gauges, and a system control unit. Schematic representation of the
system is showed in Fig. 3.
The system is being controlled manually from control panel and
evacuation process is secured by following list of incremental steps.
Vacuum system evacuation begins when forevacuum pump receives an input
start signal from system control panel and after valves opening provides
base pressure of nominally 2x[10.sup.-2] Torr. As second step a
turbomolecular pump must be enabled, when it accelerates and reaches
necessary rotational speed. In the meantime the gate valve should be
opened. After the gate is opened pumping provides pressure inside vacuum
chamber 2x[10.sup.-6] Torr. Gas system consists of two separate lines
for argon and oxygen flow with manual valves and mass flow controllers
in order to control gas flow during experiments. Gas flow can be
operated in range from 1 sccm to 10 sccm.
One of the most important parameters of vacuum system is the design
of vacuum chamber. Manufactured vacuum system is capable of providing
the lowest pressure of 8x[10.sup.-7] Torr. Given pressure can be
achieved in short period of time comparing to the coating vacuum systems
with high volume for manufacturing purpose. prior art discusses same
idea of vacuum calibration system design, where the easiest way to
obtain a constant pressure and maintain gas purity in the high vacuum
range is to pass a steady flow of the calibration gas in to the vacuum
system and out through the high vacuum pump [7]. Given the importance of
uniform pressures and the number of variables that can influence this,
it is important to test if the reference gauges and test gauges are
sensing the same pressure [7]. Design and the size of verification
chamber determine how man vacuum gauges can be mounted simultaneously.
Small size vacuum chamber provides faster system evacuation, facilitates
to maintain unchanged pressure and significantly reduces the required
volume of gas needed.
Therefore, several connection ports are necessary to perform vacuum
gauges pressure measurement comparison actions, adjustment and
calibration if reference standard is available. In order to reach
selected criteria the vacuum chamber was designed and manufactured using
mechanical machining and blacksmithing work. All the vacuum system
components are showed at fig. 4.
As illustarted there are four side covers and one cover at the top
of the chamber for vacuum gauges connections. Main idea was to design
relatively large connection ports with covers, place them symmetrically
to each other. Such design allows to connect up to four vacuum gauges on
one cover, thus a distance from pumping orifice and gas inlet port is
equal, which provides reliable vacuum gauge pressure measurements. This
way both vacuum gauges will sense identical pressure in vacuum chamber.
For the experiments there were produced several covers, without
connections ports and with different connection ports location. unique
design of vacuum chamber lets switch vacuum gauges positions, connect
several gauges simultaneously, change position of the gas inlet system.
All the manipulations made with listed elements can provide useful
information about vacuum gauge performance and pressure measurement
accuracy. After preliminary analysis of collected data, it is possible
to find relationships between pressure measurements readings, their
positioning and gas flow impact.
The system uses turbomolecular pump VARIAN Turbo-V 3K-T (5)
connected to a vacuum chamber (1) though a gate valve (4) and having a
pumping speed up to 1900 l/s for argon without inlet screen. This pump
consists of high frequency motor driving a turbine fitted with 10 bladed
stages with nominal rotational speed up to 31800 rmp, thus a water-
cooling system (10) is essential. Pumping line consists of
turbomolecular pump (5) connected to a pipe (12) through solenoid valve
and forevacuum pump connected to vacuum chamber through a pipe (11) and
through a solenoid valve (13) to vacuum chamber as can be seen at Fig.
4.
Gas system consists of several connected tubes, a manual valve (8),
a mass flow controller (7) and a four channel mass flow controller power
supply readout (9). Using power supply readout (9), a gas flow ratio can
be changed from 1 up to 10 sccm. Manual valve is a backup tool for
sealing the gas inlet line.
Mainly two types of INFICON vacuum gauges were used in vacuum
coating system and for the experiments: a diaphragm vacuum gauge CDG
025D and ionization gauges HPG 400 and BPG 402. Diaphragm vacuum gauge
is used as most reliable and stable pressure measuring instrument due to
its gas independence. It combines low relative measurement uncertainty
with large dynamic range and high stability [6]. Accuracy of gauge also
plays significant role and therefore reading uncertainty must be the
lowest, thus the diaphragm gauge could meet selected criteria with
measurement uncertainty of 0,50% from reading.
High vacuum must be reached in order to implement vacuum gauge
performance verification in broad vacuum range including its upper and
lower pressure range limits where relative measurement uncertainty
reaches the highest value. other important factor is that high pressure
provides an opportunity for vacuum gauge adjustment. Sometimes the word
"calibration" is misused to describe the process of altering
the performance of an instrument to ensure that the values it indicates
are correct within specified limits [1]. This is adjustment, but not
calibration [1]. Formally, calibration is said to be "a set of
operations the establish, under specified conditions, the relationship
between the values of quantities indicated by a measuring instrument and
the corresponding values realized by standards" [1]. Measurement
standards can be divided in two different types: primary and secondary
[3]. A primary standard is one "designed or widely acknowledged as
having the highest metrological qualities and whose value is accepted
without reference to other standards of the same quality" [1].
Secondary or transfer standards are calibrated against a primary
standard of the same unit and depend on a predictable stability to
maintain their accuracy [3]. Vacuum gauge calibration in manufacturing
conditions cannot be made due to several reasons including the reference
standard unavailability; therefore, a vacuum gauge adjustment can be
performed. Vacuum gauge operating manual describes that it's
recommended to perform a zero adjustment, when gauge is operated for the
first time and also due to long time operation or contamination, a zero
drift could occur and zero adjustment may become necessary [11]. For
adjusting the zero a gauge must be operated under the same constant
ambient conditions and in the same mounting orientation as normally
[11].
Developed vacuum system was created as a necessary equipment for
the vacuum gauges verification under manufacturing conditions in order
to test their performance using comparison method, generate adjustment,
obtain and save all the data from experiments, make analysis of
relationships between several factors and vacuum gauge pressure
measurements accuracy.
4. Experiments
In research of vacuum gauge performance, several experiments were
designed and conducted using developed verification system. As the first
step a vacuum gauge pressure measurements values can be checked by
comparison method. For the experiment it was necessary to select minimum
of two identical vacuum gauges and third one as a reference gauge. In
this particular experiment two INFICON ionization gauges HPG 400 were
selected as verifiable gauges and third gauges BPG 402 was selected as a
reference.
In INFICON HPG 400 ionization vacuum gauge data sheets stated that
the accuracy is 15% of reading in range from 10E-5 to 1 mbar [12]. Same
accuracy provided by the BPG 402 gauges in the range from 10E-8 to 10E-2
mbar [13]. After preliminary analysis of the results it was possible to
conclude that ionization gauge HPG 400 (SN 549) lacks accuracy due to
significant difference in pressure measurements comparing to reference
gauge. As an example, the first measuring point was compared for HPG 400
(SN549) and BPG 402 gauges. Difference between indicated pressures was
58% if we accept that reference gauge pressure is a true value. Prior
art discusses that in order to be able to allocate the largest possible
measuring ranges to the individual types of vacuum gauge, one has to
accept the fact that the measurement uncertainty rises very rapidly, in
some cases up to the greater extent, particularly at the upper and lower
limit [5]. First three measuring points do not exceed upper limit for
HPG 400 ionization gauges, thus accuracy at specified pressure range
complies lowest possible uncertainty. BPG type vacuum gauge accuracy at
all measuring points remains the highest due to its relative measurement
uncertainty in the middle of range limits.
In the second experiment a comparison of ionization gauges with
turned on argon correction was performed. During all experiments an
argon flow was controlled and gradually decreased. This step is
necessary to obtain a data of vacuum gauges performance with argon
correction factor. Another idea for further research is to verify, how
vacuum gauge corresponds to gas flow reduction. Ionization gauges HPG
400 (SN566) and BPG 402 graphic lines are identical and indicated
pressure values from Table 2 are identical up to eight measurement
point, which is upper range limit of HPG 400 (SN 566) where relative
measurement uncertainty rises.
Character of pressure measurements changes should be identical in
both experiments. From results it was possible to conclude the fact that
vacuum gauge performances are identical with and without gas correction
factor, but accuracy of ionization vacuum gauge HPG 400 (SN 549) exceeds
the permissible limits. As an example the sixth measuring point can be
compared for HPG 400 (SN549) and BPG 402 gauges. Difference between
reference and tested gauge at sixth point indicated 31% of deviation.
HPG 400 gauges consists of two subgauges, where Pirani gauge filament is
exposed to vacuum, thus during opening of a chamber and during
technological process, the particles are deposited on the filament.
Contamination of vacuum gauge leads to lack of accuracy and performance
characteristics.
5. Conclusion
In this paper we analysed vacuum gauge pressure deviation problem.
The vacuum system offered in this paper allows performing vacuum gauge
verification under manufacturing conditions, adjustment of vacuum in the
system, as well as allowing to decide about necessity of calibration
period, obtaining and retaining the data. Custom made system has
specific design, where vacuum gauges can be tested using various amount
of ports and locations, position of gas inlets can be switched, thus
small size chamber provides fast system evacuation reaching maximum
pressure of 8x10-7 Torr.
Ionization vacuum gauges were tested by comparison method using
custom made vacuum system. After preliminary analysis of experimental
data it was concluded that ionization vacuum gauge HPG 400 (SN549) lacks
accuracy, deviations in measurements are significant compared to
reference gauge BPG 402.
This vacuum system can be practically used in any vacuum coating
manufacturing company to verify its vacuum gauges, to adjust them, to
obtain and retain the data and also make a decision of calibration
necessity. This system has unique design which allows further
development of various approaches and methods for vacuum gauge
verification.
Further research should focus on various approaches and
experimental methods which could be implemented for vacuum gauge
verification using custom made vacuum system. In result of which a
methodology of vacuum gauge verification under manufacturing conditions
could be created.
DOI: 10.2507/27th.daaam.proceedings.089
6. References
[1] http://www.npl.co.uk, (1998). National Physical Laboratory.
Guide to the Measurement of Pressure and Vacuum, Accessed on: 25.09.2016
[2] Charles A. Bishop. (2007). Vacuum deposition onto Webs, films,
and foils, William Andrew Publishing, ISBN-13: 978-0-8155-1535-7
(978-0-8155), 13 Eaton Avenue, Norwich, NY 13815
[3] http://www.glb.nist.gov, 1992. National Institute of Standards
and Techonology. Pressure and Vacuum Measurements in Physical Methods of
Chemistry, Accessed on 25.09.2016
[4] https://www.leybold.com, 2016. Leybold. Fundamentals of Vacuum
Technology, Accessed on 25.09.2016
[5] Akram, H. M., Fasih, A. (2012). Selection criterion of gauges
for vacuum measurements of systems with diverse ranges, Physics
Procedia, Vol 32., pp. 503-512, ISSN: 1875-3892
[6] Barthelemy Daudea, Hadj Elandaloussia, b, Christof Janssena, b.
(2014). On the gas dependence of thermal transpiration and a critical
appraisal of correction methods for capacitive diaphragm gauges, Vacuum,
Vol. 104, pp. 77-87, ISSN: 0042-207X
[7] Ilknur Kocas, Gokce Sevim Sariyerli. 2012. The characterization
of new VGMS (vacuum gauge metrology system) by the traceability of multi
stage static expansion system, Measurement, Vol. 45, pp., 2430-2433,
ISSN: 0263-2241
[8] Stekleins A., Gerins E., Kromanis A. (2016). Analysis of
selection criteria for vacuum gauges and its accuracy, unpublished
[9] http://www.lesker.com, 2016. Kurt J.Lesker, 9th edition Global
Vacuum Product Guide, Accesses on: 25.09.2016
[10] http://www.noshok.com 2016, Measurement solutions, Process
Conditions That Affect Pressure Guage Accuracy [conjunction]
Performance, Accessed on: 25.092016
[11] http://inficon.com., 2016, Vacuum instruments, CDG 025D Data
Sheet, Accessed on: 26.09.2015
[12] http://inficon.com, 2016, Vacuum instruments, HPG 400 Data
Sheet, Accessed on: 26.09.2015
[13] http://inficon.com, 2016, Vacuum instruments, BPG 402 Data
Sheet, Accessed on: 26.09.2015
This Publication has to be referred as: Stekleins, A[ntons];
Gerins, E[riks] & Kromanis, A[rtis] (2016). Vacuum Gauge Performance
Verification System, Proceedings of the 27th DAAAM International
Symposium, pp.0607-0614, B. Katalinic (Ed.), Published by DAAAM
International, ISBN 978-3-902734-08-2, ISSN 1726-9679, Vienna, Austria
Caption: Fig. 1. Indicated pressure (a) without argon correction;
(b) with argon correction.
Caption: Fig. 2. A custom made vacuum verification system.
Caption: Fig. 3. Block scheme of verification vacuum system.
Caption: Fig. 4. Vacuum verification system with its components.
Caption: Fig. 5. HPG 400 and BPG 402 vacuum gauge performance
comparison.
Caption: Fig. 6. HPG 400 and BPG 402 vacuum gauge performance
comparison with argon gas.
Table 1. HPG and BPG vacuum gauges measured pressures values.
No. HPG 400 (SN 566) BPG 402 HPG 400 (SN 549)
1 2,50E-03 (Torr) 1,90E-03 (Torr) 8,00E-04 (Torr)
2 1,00E-03 (Torr) 8,00E-04 (Torr) 3,20E-04 (Torr)
3 2,10E-04 (Torr) 1,70E-04 (Torr) 6,50E-05 (Torr)
4 5,90E-05 (Torr) 5,30E-05 (Torr) 2,20E-05 (Torr)
5 3,20E-05 (Torr) 3,00E-05 (Torr) 1,40E-05 (Torr)
6 1,30E-05 (Torr) 1,40E-05 (Torr) 7,50E-06 (Torr)
7 8,20E-06 (Torr) 1,00E-05 (Torr) 5,90E-06 (Torr)
8 4,60E-06 (Torr) 7,10E-06 (Torr) 4,50E-06 (Torr)
9 1,80E-06 (Torr) 4,00E-06 (Torr) 2,70E-06 (Torr)
Table 2. HPG and BPG vacuum gauges indicated pressures
with argon correction.
No. Argon flow HPG 400 BPG 402 HPG 400
(sccm) (SN 566) (SN549)
1 10sccm 2,30E-04 2,30E-04 1,70E-04
2 9sccm 2,10E-04 2,10E-04 1,50E-04
3 8sccm 1,90E-04 1,90E-04 1,30E-04
4 7sccm 1,60E-04 1,60E-04 1,20E-04
5 6sccm 1,40E-04 1,40E-04 1,00E-04
6 5sccm 1,20E-04 1,20E-04 8,30E-05
7 4sccm 9,40E-05 9,40E-05 6,60E-05
8 3sccm 7,00E-05 7,10E-05 4,70E-05
9 2sccm 4,70E-05 4,90E-05 3,00E-05
10 1sccm 2,20E-05 2,40E-05 1,40E-05
COPYRIGHT 2017 DAAAM International Vienna
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 2017 Gale, Cengage Learning. All rights reserved.