  • 标题:Comparison of Macular Retinal Thickness among Four Optical Coherence Tomography Devices in Healthy Young Subjects
  • 本地全文:下载
  • 作者:Cho, Soo Yeon ; Hwang, Young Hoon
  • 期刊名称:Journal of the Korean Ophthalmological Society
  • 印刷版ISSN:0378-6471
  • 出版年度:2019
  • 卷号:60
  • 期号:5
  • 页码:434-439
  • DOI:10.3341/jkos.2019.60.5.434
  • 出版社:The Korean Ophthalmological Society
  • 摘要:Purpose

    This study was performed to compare macular retinal thickness measured by four different optical coherence tomography (OCT) devices.

    Methods

    Macular retinal thicknesses were measured in 32 healthy eyes of 32 subjects by using Cirrus HD-OCT (Carl Zeiss Meditec, Inc., Dublin, CA, USA), Spectralis OCT (Heidelberg engineering, Heidelberg, Germany), Topcon DRI OCT (Topcon Corp., Tokyo, Japan), and RS-3000 Advance OCT (Nidek, Co., Ltd., Gamagori, Japan). Macular retinal thickness measurements were compared among the devices.

    Results

    Average macular retinal thickness ranged 288.7 to 313.3 µm ( p < 0.01). Differences in average macular retinal thickness among the devices ranged from 0.7 to 27.3 µm; the limit of agreement (LoA) ranged from 9.7 to 24.4 µm. Among the devices, comparison between Spectralis OCT and RS-3000 Advance OCT showed the smallest difference in average macular retinal thickness (mean, 0.7 µm) and the narrowest LoA (−6.4 to 3.3 µm).

    Conclusions

    Macular retinal thickness measured by four different OCT devices showed different values. However, values obtained by Spectralis OCT and RS-3000 Advance OCT showed good agreement.

  • 关键词:Macula; Optical coherence tomography; Retina