首页    期刊浏览 2024年07月19日 星期五
登录注册

文章基本信息

  • 标题:The usefulness of CorvisST Tonometry and the Ocular Response Analyzer to assess the progression of glaucoma
  • 本地全文:下载
  • 作者:Masato Matsuura ; Kazunori Hirasawa ; Hiroshi Murata
  • 期刊名称:Scientific Reports
  • 电子版ISSN:2045-2322
  • 出版年度:2017
  • 卷号:7
  • 期号:1
  • DOI:10.1038/srep40798
  • 语种:English
  • 出版社:Springer Nature
  • 摘要:Corneal Visualization Scheimpflug Technology (CST) and Ocular Response Analyzer (ORA) measurements were carried out in 105 eyes of 69 patients with primary open-angle glaucoma. All patients had axial length (AL), central corneal thickness (CCT), intraocular pressure (IOP) with Goldmann applanation tonometry (GAT) and eight visual fields (VF)s with the Humphrey Field Analyzer. VF progression was summarized using a time trend analysis of mean total deviation (mTD) and the association between mTD progression rate and a number of ocular parameters (including CST and ORA measurements) was assessed using mixed linear regression analysis. The optimal model of VF progression selected based on the corrected Akaike Information Criteria (AICc) included ORA's corneal hysteresis (CH) parameter as well as a number of CST measurements: mTD progression rate = 1.2-0.070 * mean GAT + 0.090 * CH-1.5 * highest concavity deformation amplitude with CST + 9.4 * A1 deformation amplitude with CST-0.05 * A2 length with CST (AICc = 125.8). Eyes with corneas that experience deep indentation at the maximum deformation, shallow indentation at the first applanation and wide indentation at the second applanation in the CST measurement are more likely to experience faster rates of VF progression.
国家哲学社会科学文献中心版权所有