  • 标题:Effects of energy level, reconstruction kernel, and tube rotation time on Hounsfield units of hydroxyapatite in virtual monochromatic images obtained with dual-energy CT
  • 本地全文:下载
  • 作者:Jeong, Dae-Kyo ; Lee, Sam-Sun ; Kim, Jo-Eun
  • 期刊名称:Imaging Science in Dentistry
  • 印刷版ISSN:2233-7822
  • 出版年度:2019
  • 卷号:49
  • 期号:4
  • 页码:273-279
  • DOI:10.5624/isd.2019.49.4.273
  • 出版社:Korean Academy of Oral and Maxillofacial Radiology
  • 摘要:Purpose

    This study was performed to investigate the effects of energy level, reconstruction kernel, and tube rotation time on Hounsfield unit (HU) values of hydroxyapatite (HA) in virtual monochromatic images (VMIs) obtained with dual-energy computed tomography (DECT) (Siemens Healthineers, Erlangen, Germany). Materials and Methods

    A bone density calibration phantom with 3 HA inserts of different densities (CTWATER®; 0, 100, and 200 mg of HA/cm3) was scanned using a twin-beam DECT scanner at 120 kVp with tube rotation times of 0.5 and 1.0 seconds. The VMIs were reconstructed by changing the energy level (with options of 40 keV, 70 keV, and 140 keV). In order to investigate the impact of the reconstruction kernel, virtual monochromatic images were reconstructed after changing the kernel from body regular 40 (Br40) to head regular 40 (Hr40) in the reconstruction phase. The mean HU value was measured by placing a circular region of interests (ROIs) in the middle of each insert obtained from the VMIs. The HU values were compared with regard to energy level, reconstruction kernel, and tube rotation time. Results

    Hydroxyapatite density was strongly correlated with HU values (correlation coefficient=0.678, P <0.05). For the HA 100 and 200 inserts, HU decreased significantly at increased energy levels (correlation coefficient= −0.538, P <0.05) but increased by 70 HU when using Hr40 rather than Br40 (correlation coefficient=0.158, P <0.05). The tube rotation time did not significantly affect the HU ( P >0.05). Conclusion

    The HU values of hydroxyapatite were strongly correlated with hydroxyapatite density and energy level in VMIs obtained with DECT.

  • 关键词:Computed Tomography; Hydroxyapatite; Image reconstruction