首页    期刊浏览 2024年11月09日 星期六
登录注册

文章基本信息

  • 标题:New method of temperature and conductivity sensor calibration with improved efficiency for screening SBE41 CTD on Argo floats
  • 本地全文:下载
  • 作者:Shigeki Hosoda ; Mizue Hirano ; Takayuki Hashimukai
  • 期刊名称:Progress in Earth and Planetary Science
  • 电子版ISSN:2197-4284
  • 出版年度:2019
  • 卷号:6
  • 期号:1
  • 页码:1-25
  • DOI:10.1186/s40645-019-0310-1
  • 出版社:Springer Verlag
  • 摘要:The global ocean monitoring system would benefit from improvements in the efficiency of Conductivity, Temperature, and Depth Device (CTD) sensor screening. Here, we describe the development of a new screening system, J-Calibration, for use with the SBE41 CTD sensor on the Argo float (Sea Bird Scientific). This new system has crucial advantages over the traditional SBE-Calibration system. First, J-Calibration does not require removal of the CTD sensor unit from the Argo float body to operate in a laboratory. This feature enables technicians or operators to more efficiently check the Argo target accuracies (± 0.005 °C for temperature and ± 0.01 PSS-78 for salinity) and allows the manufacturer’s warranty to remain intact. This also allows for a more efficient basic screening system which maintains screening accuracy without the need for specialized technicians. J-Calibration reduces the screening time to 1/6th that of SBE-Calibration and does not require the preparation of large amounts of artificial seawater. J-Calibration uses 1/23rd of the volume of seawater compared to SBE-Calibration by examining calibration at only 1 temperature point (22 °C), whereas SBE-Calibration requires 7 points to achieve calibration. The J-Calibration system does require careful temperature control of the artificial seawater as it is critical to maintain a uniform water temperature throughout the experiments. To satisfy this, using a vinyl greenhouse and covering sensors with adiabatic materials and cooling packs are effective. In comparison with the accuracy and system stability of SBE-Calibration, we show that the J-Calibration system is non-inferior and therefore is suitable for use in laboratory screening prior to deployment. Based on these advantages, the J-Calibration system will make a strong contribution to the deployment of healthy Argo fleets and to the maintenance of uniform data in the global ocean.
  • 关键词:CTD sensor; Sensor screening; SBE41; Simple calibration method; Water temperature control; Global Argo observation system
国家哲学社会科学文献中心版权所有