  • 标题:Reliability and Accuracy of Six Hand-Held Blood Lactate Analysers
  • 本地全文:下载
  • 作者:Jacinta M. Bonaventura ; Jacinta M. Bonaventura ; Ken Sharpe
  • 期刊名称:Journal of Sports Science and Medicine
  • 印刷版ISSN:1303-2968
  • 出版年度:2015
  • 卷号:14
  • 期号:1
  • 页码:203-214
  • 语种:English
  • 出版社:University of Uludag
  • 摘要:The reliability and accuracy of five portable blood lactate (BLa) analysers (Lactate Pro, Lactate Pro2, Lactate Scout+, Xpress™, and Edge) and one handheld point-of-care analyser (i-STAT) were compared to a criterion (Radiometer ABL90). Two devices of each brand of analyser were assessed using 22 x 6 mL blood samples taken from five subjects at rest and during exercise who generated lactate ranging ~1-23 mM. Each sample was measured simultaneously ~6 times on each device. Reliability was assessed as the within-sample standard deviation (wsSD) of the six replicates; accuracy as the bias compared with the ABL90; and overall error (the root mean squared error (√MSE)) was calculated as the square root of (wsSD2 and bias2). The √MSE indicated that both the Edge and Xpress had low total error (~0-2 mM) for lactate concentrations <15 mM, whereas the Edge and Lactate Pro2 were the better of the portable analysers for concentrations >15 mM. In all cases, bias (negative) was the major contribution to the √MSE. In conclusion, in a clinical setting where BLa is generally <15 mM the Edge and Xpress devices are relevant, but for athlete testing where peak BLa is important for training prescription the Edge and Lactate Pro2 are preferred.
  • 关键词:Bias;precision;root mean squared error;analytical performance