出版社:Brazilian Association of Sleep and Latin American Federation of Sleep Societies
摘要:In a device based on midsagittal jaw movements analysis, we assessed a sleep-wake automatic detector as an objective method to measure sleep in healthy adults by comparison with wrist actigraphy against polysomnography (PSG). Simultaneous and synchronized in-lab PSG, wrist actigraphy and jaw movements were carried out in 38 healthy participants. Epoch by epoch analysis was realized to assess the ability to sleep-wake distinction. Sleep parameters as measured by the three devices were compared. This included three regularly reported parameters: total sleep time, sleep onset latency, and wake after sleep onset. Also, two supplementary parameters, wake during sleep period and latency time, were added to measure quiet wakefulness state. The jaw movements showed sensitivity level equal to actigraphy 96% and higher specificity level (64% and 48% respectively). The level of agreement between the two devices was high (87%). The analysis of their disagreement by discrepant resolution analysis used PSG as resolver revealed that jaw movements was right (58.9%) more often than actigraphy (41%). In sleep parameters comparison, the coefficient correlation of jaw movements was higher than actigraphy in all parameters. Moreover, its ability to distinct sleep-wake state allowed for a more effective estimation of the parameters that measured the quiet wakefulness state. Midsagittal jaw movements analysis is a reliable method to measure sleep. In healthy adults, this device proved to be superior to actigraphy in terms of estimation of all sleep parameters and distinction of sleep-wake status.
关键词:Actigraphy;Movement;Sleep;Sleep Wake Disorders