首页    期刊浏览 2024年07月06日 星期六
登录注册

文章基本信息

  • 标题:Evaluating the accuracy of hand models obtained from two 3D scanning techniques
  • 本地全文:下载
  • 作者:Fang Yu ; Lei Zeng ; Ding Pan
  • 期刊名称:Scientific Reports
  • 电子版ISSN:2045-2322
  • 出版年度:2020
  • 卷号:10
  • 期号:1
  • DOI:10.1038/s41598-020-68457-6
  • 出版社:Springer Nature
  • 摘要:The aim of this study was to identify an efficient approach for 3D imaging of hand. The 3D photographs of hand were taken with Gemini structured-light scanning system (SL scanning) and CT scanning. The 3D photographs, average time of scanning and reconstruction were compared between these two indirect techniques. The reliability, reproducibility and accuracy were evaluated in these two indirect techniques and the direct measurement (DM). Statistical differences in the measurements were assessed by 99% probability, with clinical significance at > 0.5 mm. The Gemini structured-light scanning system established a complete and smooth 3D hand photograph with shorter scanning and reconstruction time. Reproducibility of CT scanning and SL scanning methods was better (P 0.5 mm, indicating significant difference between DM and CT scanning at the level of 99% probability. In this study, the Gemini structured-light scanning system not only successfully established a complete and smooth 3D hand photograph, but also shortened the scanning and reconstruction time. Compared to the DM, measurements obtained using the two indirect techniques did not show any statistically or clinically insignificant difference in the values of the remaining 17 of 19 measurements (89.47%). Therefore, either of the two alternative techniques could be used instead of the direct measurement method.
国家哲学社会科学文献中心版权所有