期刊名称:International Journal of Advanced Computer Science and Applications(IJACSA)
印刷版ISSN:2158-107X
电子版ISSN:2156-5570
出版年度:2020
卷号:11
期号:11
DOI:10.14569/IJACSA.2020.0111165
出版社:Science and Information Society (SAI)
摘要:Multi-channel muscle arrays are commonly used as sensors in bionic prosthetic devices offering an innovative solution to recover motion in transradial amputees. This study presents preliminary assessments towards validation of a muscle armband for usage in transradial users. Analog and digital components were designed based on medical agencies’ recommendations to assess future compliance with Latin American medical device regulations. The study follows two approaches, an exploratory and pre-experimental design. Design was validated and confronted among research literature and medical device regulations. For validation, a pre-experimental design was guided by a quantitative paradigm. Muscle signal was assessed before and after the condition circuit for up to four muscle signals in real time. The present study considers both the conditioning muscle signal circuit and the embedded logic implementation to record signals from the designed muscle armband. Results show that the proposed device allows to record noninvasive signals with a frequency from 20-500 Hz.
关键词:Component; muscle armband; surface electromyography; medical device regulation; transradial users