首页    期刊浏览 2025年04月08日 星期二
登录注册

文章基本信息

  • 标题:Multi-energy spectral photon-counting computed tomography (MARS) for detection of arthroplasty implant failure
  • 本地全文:下载
  • 作者:Lawrence Chun Man Lau ; Wayne Yuk Wai Lee ; Anthony P. H. Butler
  • 期刊名称:Scientific Reports
  • 电子版ISSN:2045-2322
  • 出版年度:2021
  • 卷号:11
  • 期号:1
  • 页码:1554
  • DOI:10.1038/s41598-020-80463-2
  • 出版社:Springer Nature
  • 摘要:Abstract To determine whether state-of-the-art multi-energy spectral photon-counting computed tomography (MARS) can detect knee arthroplasty implant failure not detected by standard pre-operative imaging techniques. A total knee arthroplasty (TKA) removed from a patient was reviewed. The extracted prosthesis [NexGen Legacy Posterior Stabilized (LPS) TKA] was analyzed as were pre-operative imaging examination and compared with a MARS-CT examination obtained of the extracted TKA prosthesis. Radiographs, fluoroscopy, ultrasound and MRI preoperatively did not reveal the cause of the implant failure. MARS CT images of the extracted prosthesis clearly showed the presence of posteromedial polyethylene and tibial tray wear which is compatible with the clinical appearance of the extracted TKA. MARS can identify polyethylene insert and metallic tibial tray wear as a cause of TKA failure, that could not be identified with on standard pre-operative imaging. Although clinical MARS CT system is still under development, this case does illustrate its potential clinical usefulness. This is the first study to document how MARS CT imaging can detect orthopedic implant failure not detected by standard current imaging techniques. This system has a potential clinical application in orthopedic patients.
国家哲学社会科学文献中心版权所有