首页    期刊浏览 2024年07月05日 星期五
登录注册

文章基本信息

  • 标题:Sirius Scheimpflug–Placido versus ultrasound pachymetry for central corneal thickness: meta-analysis
  • 本地全文:下载
  • 作者:Yili Jin ; Colm McAlinden ; Yong Sun
  • 期刊名称:Eye and Vision
  • 印刷版ISSN:2326-0254
  • 出版年度:2021
  • 卷号:8
  • 期号:1
  • 页码:1-11
  • DOI:10.1186/s40662-021-00227-5
  • 出版社:BioMed Central
  • 摘要:To compare the difference in central corneal thickness (CCT) measurements in normal eyes between a rotating Scheimpflug camera combined with a Placido-disk corneal topographer (Sirius, CSO, Italy) and ultrasound pachymetry (USP). A systematic literature search was conducted for relevant studies published on PubMed, Medline, EMBASE, and the Cochrane Library and ClinicalTrials.gov from inception to August 1st, 2019. Primary outcome measures were CCT measurements between Sirius and USP. A random effects model was used to pool CCT measurements. A total of twelve studies involving 862 eyes were included in this meta-analysis. The meta-analysis found CCT measurements between Sirius and USP to be statistically significantly different (P < 0.0001). The mean difference between Sirius and USP was −11.26 μm with a 95% confidence interval (CI) (−16.92 μm, −5.60 μm). The heterogeneity was I2 = 60% (P = 0.004). CCT measurements with the Sirius Scheimpflug-Placido topographer were statistically significantly lower than USP. However, it may be argued that the mean difference of 11.26 μm is not a clinically significant difference.
  • 关键词:Meta-analysis ; Central corneal thickness ; Scheimpflug-Placido topographer ; Ultrasound Pachymetry
国家哲学社会科学文献中心版权所有