摘要:Cryoablation has been used for the treatment of various sorts of solid visceral tumors, but few are reported on gastric tumor via endoscope, in terms of accurate control of ablation site, freezing depth and effective temperature. Thus, we developed a novel device, which could perform accurate cryoablation on the stomach via endoscope. This study aimed to evaluate the efficacy and safety of the device on porcine stomach. Results showed that the novel device could provide direct view of the operation space, allowing accurate and safe ablation of the stomach. Three minutes cryoablation caused a transmural, 1 cm radius gastric lesion. On serosal side, the temperature dropped to −64.2 °C, −34.1 °C, 26.1 °C at the center, 1 cm and 2 cm from center, respectively. Histopathology revealed acute ruptured cells with damaged glands in mucosa, partial disruption in muscularis propria and serosal slight exudation. Three months later, scar formed with complete recovery of gastric structure. No active bleeding or perforation of stomach, nor injury or adhesion of adjacent organs was observed. This endoscopic cryoablation device allowed safe, full-thickness cryoablation with effective temperature, which may provide an alternative treatment for gastric tumor.