  • 标题:Comparison of limited- and large-volume cone-beam computed tomography using a small voxel size for detecting isthmuses in mandibular molars
  • 本地全文:下载
  • 作者:Elen de Souza Tolentino ; Pablo Andrés Amoroso-Silva ; Murilo Priori Alcalde
  • 期刊名称:Imaging Science in Dentistry
  • 印刷版ISSN:2233-7822
  • 出版年度:2021
  • 卷号:51
  • 页码:27-34
  • DOI:10.5624/isd.20200144
  • 出版社:Korean Academy of Oral and Maxillofacial Radiology
  • 摘要:Purpose: This study was performed to compare the ability of limited- and large-volume cone-beam computed tomography (CBCT) to display isthmuses in the apical root canals of mandibular molars. Materials and Methods: Forty human mandibular first molars with isthmuses in the apical 3 mm of mesial roots were scanned by micro-computed tomography (micro-CT), and their thickness, area, and length were recorded. The samples were examined using 2 CBCT systems, using the smallest voxels and field of view available for each device. The Mann-Whitney, Friedman, and Dunn multiple comparison tests were performed (α=0.05). Results: The 3D Accuitomo 170 and i-Cat devices detected 77.5% and 75.0% of isthmuses, respectively (P>0.05). For length measurements, there were significant differences between micro-CT and both 3D Accuitomo 170 and i-Cat(P<0.05). Conclusion: Both CBCT systems performed similarly and did not detect isthmuses in the apical third in some cases. CBCT still does not equal the performance of micro-CT in isthmus detection, but it is nonetheless a valuable tool in endodontic practice.
  • 关键词:Cone-Beam Computed Tomography; Endodontics; Molar; X-Ray Microtomography