首页    期刊浏览 2024年07月08日 星期一
登录注册

文章基本信息

  • 标题:Finite element analysis of a percutaneous aortic valve stent design
  • 本地全文:下载
  • 作者:Gideon Praveen Kumar ; Lazar Mathew
  • 期刊名称:Health
  • 印刷版ISSN:1949-4998
  • 电子版ISSN:1949-5005
  • 出版年度:2009
  • 卷号:1
  • 期号:1
  • 页码:2-7
  • DOI:10.4236/health.2009.11002
  • 出版社:Scientific Research Publishing
  • 摘要:Aim: This paper discusses the design and Finite Element Analysis (FEA) of a Percutaneous Aor-tic Valve Stent. The aim of this study was to model a percutaneous aortic valve stent and subject it to finite element analysis. The design process was carried out to meet the functional and surgical requirements. Methods and Results: Analysis was done with different materials with loads ranging from 50 kgf/mm² to 73 kgf/mm². These forces were selected because these val-ues are far greater than the normal human blood pressure which ranges from 10kPa to 16kPa. It was also to understand the mechanical behavior of different stent materials under such high pressures. A stent model was generated and its physical, mechanical and behavioral properties were studied. Finite element analysis and simulation of the model enhanced the designer to optimize the geometry suitable for perform-ance during and after implantation. The design objective for the stent is to have long term du-rability, low thrombogenicity, resistance to mi-gration and paravalvular leak. Conclusion: The analysis performed in this paper may aid in understanding the stent’s tolerable pressures ranges in comparison with the physiological pressures exerted by the heart and cardiac blood flow during abnormal cardiovascular conditions.
  • 关键词:Stent; Finite Element Analysis; Blood Flow; Aortic Valve Stenosis; Port Size
国家哲学社会科学文献中心版权所有